Research on Thermal Dosage for High Intensity Focused Ultrasound in Bovine Liver Based on the Magnetic Resonance Thermometry

机译:基于磁共振测温法的高强度聚焦超声热剂量研究

获取原文

摘要

Discussing the relationships between the simulated threshold thermal dose of coagulation and coagulation on the histological slides based on the magnetic resonance (MR) thermometry. The MR imaging guided HIFU system was used, the parameters of therapeutic transducer were: frequency-1 MHz, diameter-150 mm, focal length-150 mm. Sonications were delivered in fresh bovine liver tissue in vitro at a fixed depth of 20 mm and at varying powers. Temperature-sensitive MR images obtained during the sonications were used to estimate the temperature and the thermal dosage of each voxel in the target region was calculated. The thermal dosage of each voxel can be compared with the reference threshold thermal dosage, so as to calculate the boundary and area of the coagulated tissue. After the exposure, the tissue was dissected along the maximal section of the coagulation necrosis, and the area of the biological focal region was measured. The results from this study showed that the occurrence of tissue damage correlated well with the equivalent thermal dose calculation based on the T-mapping of MRI, and they share the same variation tendency.
机译:基于磁共振(MR)温度的组织学载玻片模拟阈值热剂量与凝固的关系。使用MR成像引导HIFU系统,治疗换能器的参数是:频率-1MHz,直径-150mm,焦距-150mm。在固定深度为20mm的固定深度和不同的功率下,在新鲜牛肝组织中递送了超声。在超声处理期间获得的温度敏感MR图像用于估计靶区域中每种体素的温度和热剂量。可以将每个体素的热剂量与参考阈值热剂量进行比较,从而计算凝固组织的边界和面积。曝光后,沿着凝血坏死的最大截面解剖组织,并测量生物焦点区域的面积。该研究的结果表明,基于MRI的T映射的等效热剂量计算,组织损伤的发生良好,它们共享相同的变异趋势。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号