首页> 外文会议>International Conference on 3D Radiation Dosimetry >Enhancement in dose sensitivity of polymer gel dosimeters composed of radiation-crosslinked gel matrix and less toxic monomers
【24h】

Enhancement in dose sensitivity of polymer gel dosimeters composed of radiation-crosslinked gel matrix and less toxic monomers

机译:高分子凝胶剂量计的剂量敏感性提高辐射交联凝胶基质和较小毒性单体的剂量敏感性

获取原文

摘要

Polymer gel dosimeters based on radiation-crosslinked hydroxypropyl cellulose gel were prepared, which comprised 2-hydroxyethyl methacrylate (HEMA) and polyethylene glycol #400 dimethacrylate (9G) as less toxic monomers and tetrakis (hydroxymethyl) phosphonium chloride (THPC) as an antioxidant. The dosimeters exposed to ~(60)Co γ-rays became cloudy at only 1 Gy. The irradiated dosimeters were optically analyzed by using a UV-vis spectrophotometer to evaluate dose response. Absorbance of the dosimeters linearly increased in the dose range from 0 to 10 Gy, in which dose sensitivity increased with increasing 9G concentration. The dose sensitivity of the dosimeters with 2 wt% HEMA and 3 wt% 9G was also enhanced by increment in THPC.
机译:制备基于放射交联的羟丙基纤维素凝胶的聚合物凝胶剂量计,其包含2-羟乙基甲基丙烯酸甲酯(HEMA)和聚乙二醇#400的二甲基丙烯酸酯(9g),为较小的毒性单体和四(羟甲基)氯化鏻(THPC)作为抗氧化剂。暴露于〜(60)COγ射线的剂量计仅为1 GY变得多云。通过使用UV-Vis分光光度计进行光学分析辐照剂量计以评估剂量反应。剂量计的吸光度在0到10Gy的剂量范围内线性增加,其中剂量敏感性随着9g浓度的增加而增加。通过THPC中的增量也增强了具有2wt%hema和3wt%9g的剂量计的剂量敏感性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号