Polymer gel dosimeters with reduced toxicity: a preliminary investigation of the NMR and optical dose-response using different monomers

机译:降低毒性的聚合物凝胶剂量计:使用不同单体对NMR和光学剂量响应的初步研究

获取原文
获取原文并翻译 | 示例
           

摘要

In this work, three new polymer gel dosimeter recipes were investigated that may be more suitable for widespread applications than polyacrylamide gel dosimeters, since the extremely toxic acrylamide has been replaced with the less harmful monomers N-isopropylacrylamide (NIPAM), diacetone acrylamide and N-vinylformamide. The new gel dosimeters studied contained gelatin (5 wt%), monomer (3 wt%), N, N'-methylene-bis-acrylamide crosslinker (3 wt%) and tetrakis (hydroxymethyl) phosphonium chloride antioxidant (10mM). The NMR response (R-2) of the dosimeters was analysed for conditions of varying dose, dose rate, time post-irradiation, and temperature during irradiation and scanning. It was shown that the dose-response behaviour of the NIPAM/Bis gel dosimeter is comparable to that of normoxic polyacrylamide gel (PAGAT) in terms of high dose-sensitivity and low dependence on dose rate and irradiation temperature, within the ranges considered. The dose response (R-2) of NIPAM/Bis appears to be linear over a greater dose range than the PAGAT gel dosimeter. The effects of time post-irradiation (temporal instability) and temperature during NMR scanning on the R-2 response were more significant for NIPAM/Bis dosimeters. Diacetone acrylamide and N-vinylformamide gel dosimeters possessed considerably lower dose-sensitivities. The optical dose-response, measured in terms of the attenuation coefficient for each polymer gel dosimeter, showed potential for the use of optical imaging techniques in future studies.
机译:在这项工作中,研究了三种新的聚合物凝胶剂量计配方,它们比聚丙烯酰胺凝胶剂量计更适合广泛应用,因为毒性极强的丙烯酰胺已被危害较小的单体N-异丙基丙烯酰胺(NIPAM),双丙酮丙烯酰胺和N-乙烯基甲酰胺。研究的新凝胶剂量计包含明胶(5 wt%),单体(3 wt%),N,N'-亚甲基双丙烯酰胺交联剂(3 wt%)和四(羟甲基)氯化phospho抗氧化剂(10mM)。分析剂量计的NMR响应(R-2),以了解剂量,剂量率,照射后时间以及照射和扫描过程中温度的变化情况。结果表明,在所考虑的范围内,NIPAM / Bis凝胶剂量计的剂量响应行为与高氧聚丙烯酰胺凝胶(PAGAT)相当,在高剂量敏感性以及对剂量率和照射温度的低依赖性方面。 NIPAM / Bis的剂量响应(R-2)在比PAGAT凝胶剂量计更大的剂量范围内呈线性关系。对于NIPAM / Bis剂量计,辐照后时间(时间不稳定)和NMR扫描过程中温度对R-2响应的影响更为显着。双丙酮丙烯酰胺和N-乙烯基甲酰胺凝胶剂量计具有较低的剂量敏感性。以每个聚合物凝胶剂量计的衰减系数来衡量的光学剂量响应显示了在未来研究中使用光学成像技术的潜力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号