Evaluation of a photon counting X-ray imaging system

机译:评估光子计数X射线成像系统

获取原文

摘要

A digital imaging system has been developed for mammography using silicon strip detectors operated in a pulse-counting mode and configured in a scanned slit geometry. Almost 100% of the photons are absorbed in the sensor, the scatter rejection is very efficient and the image formation is nearly optimal since each x-ray is processed individually. The result is a very dose-efficient system. In this paper we present measurements that verify that the performance of the read-out electronics is sufficient to count photons at, high rates with high Quantum Efficiency (QE) and a Charge Collection Efficiency (CCE) that does not limit the dose efficiency. The spatial resolution of the system was measured to provide a Modulation Transfer Function (MTF) of approximately 0.5 at a spatial frequency 10 lp/mm. Images of a mammography phantom were recorded experimentally to test overall system performance.
机译:已经为使用以脉冲计数模式操作的硅条探测器和配置在扫描的狭缝几何形状中,为乳房X线图开发了一种数字成像系统。几乎100%的光子被吸收在传感器中,散射抑制非常有效,并且图像形成几乎最佳,因为每个X射线被单独处理。结果是一个非常有效的系统。在本文中,我们提供了测量,验证读出电子器件的性能足以计算光子,高量子效率(QE)和不限制剂量效率的充电收集效率(CCE)。测量系统的空间分辨率以在空间频率10Lp / mm处提供大约0.5的调制传递函数(MTF)。实验记录乳房X线摄影幻影的图像以测试整体系统性能。