Uncertainty in the operation of the Harned Cell arising from temperature and bias potential effects

机译:从温度和偏置潜在效应产生的母线手术中的不确定性

获取原文

摘要

We have considered the influence of the variation of temperature and standard potential between cells on the results of measurements made with a Harned Cell. Our calculations indicate that when all temperature and electrode bias effects are considered, the expanded uncertainty of p(a_(H)γ_(Cl))_(o) may be as much as 40% larger than is estimated in the IUPAC guidelines.
机译:我们考虑了对细胞温度和标准电位的变化对用母线的测量结果的影响。我们的计算表明,当考虑所有温度和电极偏压效应时,P(A_(H)γ_(CL))_(O)的扩展不确定度可以高于IUPAC指南中估计的40%。