首页> 外文会议>American Society for Mass Spectrometry Conference on Mass Spectrometry and Allied Topics >DISCOVERING THE ORIGIN OF AN INTERFERING PEAK OBSERVED IN LC-MS/MS ASSAY OF KETOROLAC IN RABBIT PLASMA
【24h】

DISCOVERING THE ORIGIN OF AN INTERFERING PEAK OBSERVED IN LC-MS/MS ASSAY OF KETOROLAC IN RABBIT PLASMA

机译:在兔血浆中发现在Ketorolac的LC-MS / MS测定中观察到干扰峰的起源

获取原文

摘要

Based on the results of our investigation, polypropylene (PP) tubes were implemented for the extraction portion of the method. The assay validation with LLOQ of 0.1 ng/mL was completed successfully meeting current industry standards. Precautions have always been taken for high sensitivity assays with low LLOQ by utilizing LC/MS grade solvents, ensuring cleanliness of glassware and instrumentation, etc... Additional controls and evaluations have also been implemented in this lab to ensure the ruggedness of assays when lowering the limit of quantitation from established methods.
机译:基于我们研究的结果,为该方法的提取部分实施了聚丙烯(PP)管。与LLOQ为0.1 Ng / ml的测定验证成功完成了当前行业标准。通过利用LC / MS级溶剂,确保玻璃器皿和仪器的清洁度以及额外的控制和评估,在该实验室中也实施了预防措施,确保了额外的控制和评估,以确保降低时的测定粗糙度既定方法的定量限。