首页> 外文会议>Conference on Computers in Cardiology >Development of a Pediatric ECG Rhythm Database for the Assessment of the Rhythm Analysis Algorithms of Automated External Defibrillators
【24h】

Development of a Pediatric ECG Rhythm Database for the Assessment of the Rhythm Analysis Algorithms of Automated External Defibrillators

机译:用于评估自动外离除颤器节奏分析算法的儿科ECG节奏数据库

获取原文

摘要

In 2003 the International Liaison Committee on Resuscitation (ILCOR) approved the use of Automatic External Defibrillators (AED) in children under 8 years of age. Two independent studies, conducted with commercial AEDs on proprietary pediatric ECG databases, provided the necessary scientific evidence. We have developed a database of pediatric rhythms for the evaluation of the rhythm analysis algorithms of AEDs. We collected archived ECG studies from patients under 14 years of age in two Spanish hospitals. The extracted rhythm samples contain a single rhythm, no artefact and have a minimum duration of 5 seconds. Three independent cardiologists classified the rhythm samples according to the American Heart's Association (AHA) guidelines; a consensus decision was reached when divergences arose. We compiled a total of 674 rhythm samples from 363 patients. The cardiologists classified 112 rhythm samples from 32 patients as shockable and 552 rhythm samples from 337 patients as nonshockable; 10 rhythm samples were classified as intermediate. The database contains enough nonshockable rhythm samples to meet the AHA requirements for the assessment of the specificity of AED rhythm analysis algorithms. We are currently working to complete the database with enough shockable rhythm samples to assess the sensitivity of AED rhythm analysis algorithms.
机译:2003年,国际联合会复苏委员会(ILCOR)批准在8岁以下的儿童中使用自动外部除颤器(AED)。在专有儿科社会数据库上与商业AEDS进行的两项独立研究提供了必要的科学证据。我们开发了一种儿科节律数据库,用于评估AED的节奏分析算法。我们在两个西班牙医院的14岁以下的患者中收集了存档的心电图研究。提取的节奏样品含有单个节律,没有人工制品并且具有5秒的最小持续时间。三位独立的心脏病学家根据美国心脏协会(AHA)指导方针分类节奏样本;在分歧时达成共识决定。我们从363名患者中编制了674个节奏样本。心脏病学家从32名患者中分类了112个节奏样本,从337名患者那里为非可接受的患者提供了552名患者的样品;将10个节奏样品归类为中间体。该数据库包含足够的非值可易应用的节奏样本,以满足评估AED节奏分析算法的特异性的AHA要求。我们目前正在努力完成具有足够可震性节律样本的数据库,以评估AED节奏分析算法的灵敏度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号