首页> 外文会议>Annual International Conference on Micro Electro Mechanical Systems >INTERGRATED RESONANT CANTILEVER SENSORS WITH 2nd TORSION-MODE FOR fg-LEVEL RESOLUBLE BIOMEDICAL DETECTION
【24h】

INTERGRATED RESONANT CANTILEVER SENSORS WITH 2nd TORSION-MODE FOR fg-LEVEL RESOLUBLE BIOMEDICAL DETECTION

机译:集成谐振悬臂传感器,具有用于FG级分离生物医学检测的第二扭转模式

获取原文

摘要

Presented is a silicon microcantilever resonant sensor with Lorentz-force exciting and piezoresistive sensing elements integrated for ultra-resoluble biomedical detection. A 2nd torsional resonance mode is excited for high Q-factor, high mass sensitivity and high mass resolution. After the sensor design and fabrication, the sensing properties are tested by biotin-avidin specific binding. With ng/mL level immuno-sensing experiment for the hepatoma antigene of AFP, the sensor shows promise for cancer early-stage prognosis.
机译:提出是一种硅片微膜谐振传感器,具有Lorentz-Force励磁和压阻式传感元件,集成用于超分子生物医学检测。为高Q系数,高质量敏感性和高质量分辨率激发第二扭转共振模式。在传感器设计和制造之后,通过生物素 - 抗生物素蛋白特异性结合测试感测性质。随着AFP的肝癌抗原的Ng / mL水平免疫感测实验,传感器显示了癌症早期预后的承诺。