Pulse wave measurement in human using piezoresistive cantilever on liquid

机译:用压阻悬臂在液体中使用压阻悬臂的脉搏波测量

获取原文

摘要

We propose a device that can measure pulse waves at various points on human body with high sensitivity. Pulse wave velocity was calculated from a synchronized pulse wave measurement on two points. The device had a piezoresistive cantilever placed on silicone oil. The cantilever with oil was embedded in polydimethylsiloxane (PDMS). Pressure waves from arteries can be well conveyed to the cantilever, for the human-skin-like acoustic impedance of the silicone oil and PDMS. The signal to noise ratio of the device was ~80 dB in 10-100 Hz, when excited ~1 μm in displacement.
机译:我们提出了一种能够在人体上具有高灵敏度的各个点的脉搏的装置。从两点上的同步脉冲波测量计算脉冲波速度。该装置有一个压阻悬臂放在硅油上。将悬臂嵌入聚二甲基硅氧烷(PDMS)中。来自动脉的压力波可以良好地输送到悬臂,用于硅油和PDMS的人皮肤样声阻抗。当位移时激发〜1μm时,器件的信噪比为10-100 Hz。