首页> 外文会议>Design and quality for biomedical technologies III >High count rate pseudo-random single photon counting system for time-resolved diffuse optical imaging
【24h】

High count rate pseudo-random single photon counting system for time-resolved diffuse optical imaging

机译:高计数率伪随机单光子计数系统,用于时间分辨漫射光学成像

获取原文
获取原文并翻译 | 示例

摘要

Pseudo-random single photon counting (PRSPC) is a new time-resolved method combining the spread spectrum time-resolved optical measurement method with single photon counting. A continuous wave laser diode is modulated with a pseudo-random bit sequence, while a single photon detector is used to record the pulse sequence in response to the modulated excitation. Periodic cross-correlation is performed to retrieve the impulse response. Compared with conventional time-correlated single photon counting (TCSPC), PRSPC enjoys many advantages such as low cost and high count rate without compromising the sensitivity and time-resolution. In this paper, we report a PRSPC system that can be used for high speed acquisition of the temporal spread function of diffuse photons. It can reach a photon count rate as high as 3Mcps (counts per second). Experimental work has been conducted to demonstrate the system performance.
机译:伪随机单光子计数(PRSPC)是一种新的时间分辨方法,将扩展频谱时间分辨光学测量方法与单光子计数相结合。连续波激光二极管采用伪随机位序列进行调制,而单个光子检测器用于响应调制后的激励来记录脉冲序列。执行周期性互相关以检索脉冲响应。与传统的时间相关单光子计数(TCSPC)相比,PRSPC具有许多优势,例如低成本和高计数率,而不会影响灵敏度和时间分辨率。在本文中,我们报告了一种PRSPC系统,该系统可用于高速获取扩散光子的时间扩散函数。它可以达到高达3Mcps(每秒计数)的光子计数率。已经进行了实验工作以证明系统性能。