首页> 外文会议>Coherence Domain Optical Methods and Optical Coherence Tomography in Biomedicine X; Progress in Biomedical Optics and Imaging; vol.7 no.2 >Three dimensional investigation of in vivo anterior eye segments by swept-source optical coherence tomography with ready-for-shipping scanning light source
【24h】

Three dimensional investigation of in vivo anterior eye segments by swept-source optical coherence tomography with ready-for-shipping scanning light source

机译:扫描源光学相干断层扫描和现成的扫描光源对体内前眼节的三维研究

获取原文
获取原文并翻译 | 示例

摘要

A two- and three- dimensional swept source optical coherence tomography (SS-OCT) system which uses a ready-to-ship scanning light source is demonstrated. The light source has the center wavelength of 1.31 μm, the -3 dB wavelength range of 110 nm, the scanning rate of 20 KHz and high linearity of frequency scanning. A simple calibration method using a fringe analysis technique for spectral rescaling is presented. This SS-OCT is capable of realtime display of two-dimensional OCT, and can take three-dimensional OCT with the measurement time of 2 s. In vivo human anterior eye segments are investigated both two- and three- dimensionally. The system sensitivity is experimentally determined as 113 dB.
机译:演示了使用现成的扫描光源的二维和三维扫频光源光学相干断层扫描(SS-OCT)系统。光源的中心波长为1.31μm,-3 dB波长范围为110 nm,扫描速率为20 KHz,并且具有高线性度的频率扫描。提出了一种使用条纹分析技术进行光谱重新缩放的简单校准方法。该SS-OCT能够实时显示二维OCT,并且可以以2 s的测量时间进行三维OCT。在体内对人的前眼节进行了二维和三维研究。实验确定系统灵敏度为113 dB。