Dose reduction in orthodontic lateral cephalography: dosimetric evaluation of a novel cephalographic thyroid protector (CTP) and anatomical cranial collimation (ACC)

机译:正畸外侧头颅造影术中的剂量减少:新型头颅甲状腺造影术(CTP)和颅骨准直解剖学(ACC)的剂量学评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objectives: To test the dose-reducing capabilities of a novel thyroid protection device and a recently introduced cranial collimator to be used in orthodontic lateral cephalography. Methods: Cephalographic thyroid protector (CTP) was designed to shield the thyroid while leaving the cervical vertebrae depicted. Using a RANDO® head phantom (The Phantom Laboratory, Salem, NY) equipped with dosemeters and a Proline XC (Planmeca, Helsinki, Finland) cephalograph, lateral cephalograms were taken, and the effective dose (ED) was calculated for four protocols: (1) without shielding; (2) with CTP; (3) with CTP and anatomical cranial collimator (ACC); and (4) with a thyroid collar (TC). Results: The ED for the respective protocols was (1) 8.51; (2) 5.39; (3) 3.50; and (4) 4.97 µSv. The organ dose for the thyroid was reduced from 30.17 to 4.50 µSv in Protocols 2 and 3 and to 3.33 µSv in Protocol 4. Conclusions: The use of just the CTP (Protocol 2) resulted in a 36.8% reduction of the ED of a lateral cephalogram. This was comparable to the classical TC (Protocol 4). A 58.8% reduction of the ED was obtained when combining CTP and ACC (Protocol 3). The dose to the radiosensitive thyroid gland was reduced by 85% in Protocols 2 and 3 and by 89% in Protocol 4.
机译:目的:测试用于正畸外侧头颅成像的新型甲状腺保护装置和最近推出的颅骨准直仪的减量能力。方法:设计颅骨图甲状腺保护器(CTP)以保护甲状腺,同时保留所描绘的颈椎骨。使用配备剂量计的RANDO®头部幻影(纽约州塞勒姆幻影实验室)和Proline XC(芬兰芬兰赫尔辛基的普兰梅卡)进行头颅侧位图拍摄,并针对以下四个方案计算有效剂量(ED):( 1)无屏蔽; (2)具有CTP; (3)具有CTP和解剖颅准直仪(ACC); (4)有甲状腺项圈(TC)。结果:相应方案的ED为(1)8.51; (2)5.39; (3)3.50; (4)4.97μSv。甲状腺的器官剂量在方案2和方案3中从30.17降低到4.50µSv,在方案4中降低到3.33µSv。结论:仅使用CTP(方案2)可以使外侧ED降低36.8%。头颅图。这与经典的TC(协议4)相当。当将CTP和ACC结合使用时(方案3),ED降低了58.8%。方案2和方案3降低了对放射敏感性甲状腺的剂量,方案4降低了89%。

著录项

相似文献

  • 外文文献
  • 中文文献
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号