Repeatability and Reproducibility of Compression Strength Measurements Conducted According to ASTM E9

机译:根据ASTM E9进行的抗压强度测量的可重复性和可重复性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Ten commercial laboratories participated in an interlaboratory study to establish the repeatability and reproducibility of compression strength tests conducted according to ASTM International Standard Test Method E9. The test employed a cylindrical aluminum AA2024-T351 test specimen. Participants measured elastic modulus and 0.2 % offset yield strength, YS(0.2 % offset), using an extensometer attached to the specimen. The repeatability and reproducibility of the yield strength measurement, expressed as coefficient of variations were cv(sub r)= 0.011 and cv(sub R)= 0.020 The reproducibility of the test across the laboratories was among the best that has been reported for uniaxial tests. The reported data indicated that using diametrically opposed extensometers, instead of a single extensometer doubled the precision of the test method. Laboratories that did not lubricate the ends of the specimen measured yield stresses and elastic moduli that were smaller than those measured in laboratories that lubricated the specimen ends. A finite element analysis of the test specimen deformation for frictionless and perfect friction could not explain the discrepancy, however. The modulus measured from stress-strain data were reanalyzed using a technique that finds the optimal fit range, and applies several quality checks to the data. The error in modulus measurements from stress-strain curves generally increased as the fit range decreased to less than 40 % of the stress range.
机译:十个商业实验室参加了一项实验室间研究,以建立根据ASTM国际标准测试方法E9进行的抗压强度测试的可重复性和可再现性。该测试采用了圆柱形铝AA2024-T351测试样品。参与者使用附接到样品的引伸计测量弹性模量和0.2%偏移屈服强度YS(0.2%偏移)。屈服强度测量的可重复性和可重复性,用变异系数表示为cv(sub r)= 0.011和cv(sub R)= 0.020。整个实验室的测试可重复性是单轴测试中最好的。 。报告的数据表明,使用完全相反的引伸计,而不是单个引伸计,可使测试方法的精度提高了一倍。未润滑样品末端的实验室测得的屈服应力和弹性模量小于在润滑样品末端的实验室中测得的屈服应力和弹性模量。然而,对无摩擦和完美摩擦的试样变形的有限元分析不能解释这一差异。使用找到最佳拟合范围的技术对从应力-应变数据测得的模量进行了重新分析,并对数据进行了多次质量检查。当拟合范围减小到小于应力范围的40%时,应力-应变曲线的模量测量误差通常会增加。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号