首页> 美国政府科技报告 >Ultrasonic and Optical Evaluation of Surgical Implant Materials and Devices. A Durability Study of Pericardial Bioprostheses
【24h】

Ultrasonic and Optical Evaluation of Surgical Implant Materials and Devices. A Durability Study of Pericardial Bioprostheses

机译:外科植入材料和器械的超声和光学评估。心包生物假体的耐久性研究

获取原文

摘要

Laser Doppler Anemometry (LDA) and accelerated fatigue testing were used in an attempt to assess the durability of two cardiac value bioprostheses. The LDA system was used to monitor the function of the cardiac valves over time. This was done through flow characterization in an aortic flow chamber, designed to closely simulate in vivo conditions, both in the near vicinity (sinuses of valsalva region) and also somewhat downstream (aortic region) from the values. The accelerated fatigue tester was operated by opening and closing the valves at a rate of 1300 R.P.M., about 18 x the normal rate. The results from the two test valves indicate a definite change in the flow characteristics downstream from the valve after certain accelerated test intervals. The high velocity cross-sectional flow area seems to increase over time in use, causing a decrease in the peak velocity. The tissue became more flaccid in certain areas, and tears were apparent at about 9.4 million cycles for the Ionescu-Shiley valve and at 24 million cycles for the Carpentier-Edwards valve. The use of Doppler ultrasound as a technique for monitoring the function of bioprostheses over time in vivo is also discussed.

著录项

相似文献