A comparison between the Phase Doppler Particle Analyzer and the combined measurements from Particle Measuring Systems' Forward Scattering Spectrometer Probe and the Optical Array Probe was conducted in an icing wind tunnel using NASA Icing Research Tunnel spray nozzles to produce the supercooled water droplet cloud. Clouds having a range of volume median diameters from 10 to greater than 50 microns were used for the instrument comparisons. A volume median diameter was calculated from combining the droplet distributions of the Optical Array Probe and the Forward Scattering Spectrometer Probe. A comparison of the combined volume median diameters and the Phase Doppler Particle Analyzer volume median diameters showed agreement from 10 microns up to 30 microns. Typical drop size distributions from the Phase Doppler Particle Analyzer, the Forward Scattering Spectrometer Probe, and Optical Array Probe are presented for several median volume diameters. A comparison of the distributions illustrates regions of the distributions where there is good agreement and other regions where there are discrepancies between the Phase Doppler Particle Analyzer and the Particle Measuring Systems' droplet size instruments.