首页> 美国政府科技报告 >Performance Testing of Radiobioassay Laboratories: In Vitro Measurements (Urinalysis): Final Report
【24h】

Performance Testing of Radiobioassay Laboratories: In Vitro Measurements (Urinalysis): Final Report

机译:Radiobioassay实验室的性能测试:体外测量(尿液分析):最终报告

获取原文

摘要

The report provides results of the two-round nationwide in vitro bioassay intercomparison study. Conclusions were based on analyses by 35 bioassay laboratories of nearly 1400 artificial urine samples containing known quantities of radionuclides. The test radionuclides were H, sup 89 Sr, sup 238 Pu, sup 241 Am, sup 137 Cs, sup 60 Co, and natural uranium. The data reported included background count rates, total samples counts, counting times, counting efficiencies, sample yields, and estimated errrors of the determinations. The measurement data were evaluated according to statistical methods presented in the November 1985 version of the draft ANSI Standard N13.30. If a laboratory failed a performance test for any one of the three criteria, the laboratory was considered to have failed the test for that category. 22 refs., 18 figs., 14 tabs. (ERA citation 13:031251)