Field and Laboratory Application of a Gas Chromatograph Low Thermal Mass Resistively Heated Column System in Detecting Traditional and Non- Traditional Chemical Warfare Agents Using Solid Phase Micro-Extraction

机译:气相色谱低热质电阻加热柱系统在固相微萃取检测传统和非传统化学战剂中的现场和实验室应用

获取原文

摘要

Emergency response units (ERUs) need a chemical sampling and analysis method that will allow rapid, and accurate on-scene detection and identification of dangerous chemicals. Fast gas chromatography (FGC) using a low thermal mass, resistively heated, toroidal (LTM-RHT) column, along with solid phase microextraction (SPME) was used as an analytical method to sample and detect frank contamination of traditional chemical warfare agents (CWAs), CWA precursors, and non-traditional CWAs (i.e. narcotics). The research results were used to evaluate the ability of the LTM-RHT column to effectively separate and identify a complex mixture of CWAs in the laboratory and in the field, and to evaluate the ability of the LTM-RHT column coupled to a mass spectrometer to rapidly separate and identify a complex mixture of narcotics. Various sampling times and column temperature ramping rates were used to evaluate peak resolution (R), column efficiency (CE) and/or peak retention time (RT). The results of the CWA sampling and analysis indicate the LTM-RHT column assemblies could be used by an (ERU) to effectively analyze and detect frank chemical warfare agent contaminated clothing up to 75% faster than a column that is heated using a standard ramping rate for an air bath oven (ABO). The decrease in analysis time, to include system cool-down time, will allow for a potentially significant increase in the number of samples that can be analyzed within a given time period. The data of the CWA precursors shows that identification with sufficient peak separation of volatile analytes via FGC is obtainable. The work with non-traditional CWAs demonstrated that FGC can significantly reduce analysis retention time while maintaining adequate peak resolution, and column efficiency.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号