This paper presents a technique tested in vitro and in animals and used in humans to measure carotid artery blood flow continuously by placing a transducer over the skin near the artery. The transducer consists of a pulse echo crystal used to locate the carotid artery and measure its diameter and wall velocity. It also has a pair of Doppler shift velocity crystals used to measure the velocity of the artery wall and the blood flow in the vessel. It was found that the pulsatile artery wall distention was between 8% and 13% of the diastolic diameter. The peak blood flow varied inversely with pulse rate. Occlusion of one carotid resulted in an increase in diastolic but not systolic flow in the other. (Author)