首页> 美国政府科技报告 >Flow-Injection Analysis Utilizing a Spectrally Segmented Photodiode-Array Inductively Coupled Plasma Emission Spectrometer I. Microcolumn Preconcentration for the Determination of Molybdenum
【24h】

Flow-Injection Analysis Utilizing a Spectrally Segmented Photodiode-Array Inductively Coupled Plasma Emission Spectrometer I. Microcolumn Preconcentration for the Determination of Molybdenum

机译:利用光谱分段光电二极管阵列电感耦合等离子体发射光谱仪进行流动注射分析I.微柱预富集测定钼

获取原文

摘要

A flow injection analysis (FIA) system incorporating a microcolumn of activated alumina was used for the determination of molybdenum by inductively coupled plasma-atomic emission spectrometry (ICP-AES). The transient signal resulting from the injection of molybdenum was determined by a spectrally segmented photodiode-array spectrometer. Linear calibration was established for molybdenum over a concentration range 0 - 1000 ng/ml. The detection capability was approximately 50 times better than offered by comparable continuous-nebulization systems. The detection limit for molybdenum was 0.2 ng/ml, and the relative standard deviations at 500 and 10 ng/ml were about 2 and 5%, respectively. The ability of the multichannel photodiode array spectrometer to correct for background shifts which occur during signal integration period provided improved precision and accuracy. Application to the analysis of molybdenum in standard sea water (NASS-2) was demonstrated. Flow injection analysis, Preconcentration, Inductively coupled plasma, Multielement analysis, Photodiode array spectrometer. (mjm)