首页> 美国政府科技报告 >Differentially Pumped Electron-Energy-Loss Spectrometer with MultichannelDetector for Time-Resolved Studies at Intermediate Ambient Pressures
【24h】

Differentially Pumped Electron-Energy-Loss Spectrometer with MultichannelDetector for Time-Resolved Studies at Intermediate Ambient Pressures

机译:具有多通道检测器的差分泵浦电子能量损失光谱仪,用于中间环境压力下的时间分辨研究

获取原文

摘要

The design, construction, and operation of a high-resolution electron-energy-lossspectrometer featuring a position-sensitive resistive anode detector and differential pumping are described. The position-sensitive detector provides a 66-fold increase in acquisition speed over a single-channel spectrometer. The differential pumping provides pressure isolation of 105 between the spectrometer elements, except the lenses, and the sample region. An optically isolated multicomputer-based control and data acquisition system provides complete spectrometer control and automated sample cleaning and characterization. The spectrometer, together with a differentially pumped supersonic molecular beamline, permits in situ measurements of the interaction of a wide range of gas molecules with solid surfaces.