首页> 外文期刊>Physics in medicine and biology. >A systematic review of the precision and accuracy of dose measurements in photon radiotherapy using polymer and Fricke MRI gel dosimetry
【24h】

A systematic review of the precision and accuracy of dose measurements in photon radiotherapy using polymer and Fricke MRI gel dosimetry

机译:对使用聚合物和Fricke MRI凝胶剂量法的光子放射治疗中剂量测量的准确性和准确性的系统评价

获取原文
获取原文并翻译 | 示例

摘要

The purpose of this work is to undertake a critical appraisal of the evidence in the published literature concerning the basic parameters of accuracy and precision associated with the use of Fricke and polymer gels (in conjunction with MR imaging) as radiation dosimeters in photon radiotherapy, condensing and analysing the body of published information (to the end of April 2002). A systematic review was undertaken addressing specific issues of precision and accuracy asking defined questions of the published literature. Accuracy and precision in relation to gel dosimetry were defined. Information was obtained from published, peer-reviewed journals. A defined search strategy utilizing MeSH headings and keywords, with extensive use of cross-referencing, identified 115 references dealing with gel dosimetry. Exclusion criteria were used to select only data from publications which would give unequivocal evidence. For accuracy, results had to be compared with an ionization chamber as gold standard and all gel samples had to be manufactured in the same batch. For precision, in addition to gels being from the same batch, samples must all have been irradiated at the same and scanned simultaneously (or within a short time frame). Many results were found demonstrating 'dose mapping' examples using gels. However, there were very few publications containing firm evidence of precision and accuracy. There was no evidence which fulfilled our criteria about accuracy or precision using Fricke gels. For polymer gels only one paper was found for accuracy 4% (Low et al 1999 Med. Phys. 26 1542-51)) and precision (1.7% (Baldock et al 1998 Phys. Med. Biol. 43 695-702)); however, both were carried out at only one dose level. If the exclusion criteria were relaxed to include accuracy results comparing gel to a non gold standard dosimeter (e.g. TLD), results give a median accuracy of 10% (range 8-23.5%) for polymer gel (Cosgrove et al 2000 Phys. Med. Biol. 45 1195-210, De Deene et al 1998 Radiother. Oncol. 48 283-91, Farajollahi et al 2000 Br. J. Radiol. 72 1085-92, McJury et al 1999b Phys. Med. Biol. 44 2431-44, Murphy et al 2000b Phys. Med. Biol. 45 835-45, Oldham et al 2001 Med. Phys. 28 1436-45) and 5% for Fricke gel (Chan and Ayyangar 1995b Med. Phys. 22 1171-5). Evidence also points to accuracy worsening at lower dose levels for both gels. The precision data should be viewed with caution as repeated MR measurements were not performed with the same samples. The only precision data for Fricke gels was 1.5% (Johansson Bask et al 1998 Phys. Med. Biol. 43 261-76), but for zero dose. In conclusion, despite the amount of published data, sparse research has been undertaken which provides clear evidence of the accuracy and precision for both gels. That which has been published has used higher doses than would be routine in radiotherapy. The basic radiation dosimeter qualities of accuracy and precision have yet to be fully quantified for polymer and Fricke gels at clinically relevant dose levels.
机译:这项工作的目的是对公开文献中有关与使用Fricke和聚合物凝胶(结合MR成像)作为光子放射治疗中的辐射剂量计相关的准确性和精密度的基本参数的证据进行严格的评估,并分析已发布信息的主体(至2002年4月底)。进行了系统的审查,以解决精确度和准确性的特定问题,并询问已发表文献的明确问题。定义了与剂量学有关的准确性和精确性。信息来自出版的,经过同行评审的期刊。使用MeSH标题和关键字的定义搜索策略,以及交叉引用的广泛使用,确定了115个涉及凝胶剂量的引用。排除标准仅用于从出版物中选择能够提供明确证据的数据。为了准确性,必须将结果与电离室作为金标准进行比较,并且所有凝胶样品必须在同一批次中生产。为了精确起见,除了来自同一批次的凝胶外,还必须所有样品都在同一时间照射并同时(或在短时间内)扫描。发现了许多结果,证明了使用凝胶的“剂量映射”实例。但是,很少有出版物包含精确性和准确性的确凿证据。没有证据能够证明我们使用Fricke凝胶的准确性或精密度。对于聚合物凝胶,仅发现一张纸的准确度为4%(Low等人,1999 Med。Phys.26 1542-51)和精确度(1.7%(Baldock等人,1998 Phys.Med.Biol.43695-702))。然而,两者都仅以一种剂量水平进行。如果放宽排除标准,以包括将凝胶与非金标准剂量计(例如TLD)进行比较的准确性结果,则结果表明聚合物凝胶的中值准确性为10%(范围8-23.5%)(Cosgrove等2000 Phys。Med。 Biol.45 1195-210,De Deene等1998 Radiother.Oncol.48 283-91,Farajollahi等2000 Br.J.Radiol.72 1085-92,McJury等1999b Phys.Med.Biol.44 2431-44 ,Murphy等人2000b Phys。Med。Biol。45 835-45,Oldham等人2001 Med。Phys。28 1436-45)和Fricke凝胶的5%(Chan and Ayyangar 1995b Med。Phys。22 1171-5)。证据还表明,两种凝胶在较低剂量水平下的准确性都会下降。应当仔细查看精度数据,因为没有对相同样品进行重复MR测量。 Fricke凝胶的唯一精密度数据为1.5%(Johansson Bask等1998 Phys。Med。Biol。43 261-76),但剂量为零。总之,尽管有公开的数据,但已经进行了稀疏的研究,这为两种凝胶的准确性和精确性提供了清晰的证据。已经发表的剂量比放射治疗中的常规剂量使用了更高的剂量。在临床相关剂量水平下,聚合物和Fricke凝胶的基本辐射剂量计质量的准确性和精密度尚待充分量化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号