A 62-year-old man with a history of familial nonischemic dilated cardiomyopathy, sick sinus syndrome, atrial fibrillation, and complete heart block was implanted with a St. Jude Atlas-II HF biventricular implantable cardioverter defibrilla-tor (BiV ICD; St. Jude Medical, St. Paul, MN, USA), with a Guidant 0158 integrated bipolar ventricular lead (Guidant Corp., St. Paul, MN, USA) and a Medtronic 5554 bipolar atrial lead (Medtronic Inc., Minneapolis, MN, USA) in 2006. He had a HeartMate II left ventricular assist device (LVAD; Thoratec Inc., Pleasanton, CA, USA) implantation in March 2009, as bridge to cardiac transplantation for end-stage heart failure. His BiV ICD was programed as follows: DDDR mode with a base rate of 85 beats/min and a maximum track rate of 120 beats per minute (bpm) with paced and sensed atrioventricular delays being 190 ms and 150 ms, respectively. His ventricular noise reversion mode was VOO and atrial tachycardia detection rate was 180 bpm. The "tachy" parameters of the device were programed as follows: ventricular tachycardia (VT)-l = 440 ms, VT-2 = 315 ms, ventricular fibrillation (VF) = 270 ms, and all zones had to fulfill 12 intervals.
机译:一位具有家族性非缺血性扩张型心肌病,病态窦房结综合征,房颤和完全性心脏传导阻滞病史的62岁男人被植入了St.Jude Atlas-II HF双心室可植入式心脏复律除颤器(BiV ICD; Jude Medical,美国明尼苏达州圣保罗),带有Guidant 0158集成双极心室导线(Guidant Corp.,美国明尼苏达州圣保罗)和Medtronic 5554双极心房导线(Medtronic Inc.,美国明尼苏达州明尼阿波利斯) )于2006年。他于2009年3月植入了HeartMate II左心室辅助装置(LVAD;美国加利福尼亚州普莱森顿的Thoratec Inc.),作为通向终末期心力衰竭的心脏移植的桥梁。他的BiV ICD编程如下:DDDR模式,基本速率为85次/分钟,最大跟踪速率为120次/分钟(bpm),带节奏的和感测到的房室延迟分别为190 ms和150 ms。他的心室消音模式为VOO,房性心动过速检出率为180 bpm。该设备的“心动过速”参数的编程如下:室性心动过速(VT)-1 = 440毫秒,VT-2 = 315毫秒,心室纤颤(VF)= 270毫秒,所有区域必须满足12个间隔。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号