首页> 外文期刊>Skeletal radiology >Precision and intersite correlation of bone densitometry at the radius, tibia and femur with peripheral quantitative CT.
【24h】

Precision and intersite correlation of bone densitometry at the radius, tibia and femur with peripheral quantitative CT.

机译:骨密度测定法在,骨,胫骨和股骨处的精确度和部位间相关性与外周定量CT。

获取原文
获取原文并翻译 | 示例

摘要

OBJECTIVE: To compare the in situ precision of peripheral quantitative CT (pQCT) at the radius, tibia and femur, and to analyze the intersite correlation, in order to determine whether measurements at the lower extremity reproduce results at the radius or are of additional informative value. DESIGN AND MATERIAL: pQCT measurements were performed in 86 elderly cadavers (mean age 80.5 years) at trabecular and cortical locations in the radius, tibia and femur, determining densitometric (bone mineral content and density) as well as geometric parameters (cross-sectional area, cortical thickness, polar moment of inertia and others). In 14 cadavers, repeated measurements were obtained at all sites on four different days. RESULTS AND CONCLUSIONS: At cortical sites, the precision for the densitometric and geometric variables ranged from 0.4% to 4.3%, and was similar for the radius, tibia and femur. At trabecular locations, the reproducibility of the density measurements ranged from 1.8% to 2.5% at the radius, and from 3.2% to 5.9% at the femur and tibia. The intersite correlation of the total bone mineral content ranged from 0.87 and 0.97 at cortical sites, and from 0.63 to 0.85 at trabecular locations. The trabecular density showed a higher similarity between the tibia and femur (r=0.68-0.78) than between the radius and the lower extremity (r=0.41-0.45). The results demonstrate a substantial heterogeneity of trabecular bone in elderly individuals and advocate measurements directly at the site of clinical or scientific interest.
机译:目的:比较peripheral骨,胫骨和股骨周围定量CT(pQCT)的原位精度,并分析部位间的相关性,以确定下肢的测量结果是否在the骨处再现结果或具有更多信息价值值。设计和材料:在Q骨,胫骨和股骨的小梁和皮质位置,对86名老年尸体(平均年龄80.5岁)进行了pQCT测量,确定了光密度(骨矿物质含量和密度)以及几何参数(横截面积) ,皮质厚度,极地惯性矩等)。在14个尸体中,在四个不同的日期在所有位置进行了重复测量。结果与结论:在皮层部位,光密度和几何变量的精度范围为0.4%至4.3%,,骨,胫骨和股骨的精度相似。在小梁位置,密度测量的可重复性在半径范围内为1.8%至2.5%,在股骨和胫骨范围内为3.2%至5.9%。总骨矿物质含量的部位间相关性在皮质部位为0.87至0.97,在小梁部位为0.63至0.85。小梁密度显示胫骨和股骨之间的相似性更高(r = 0.68-0.78),而the骨和下肢之间的相似性更高(r = 0.41-0.45)。结果表明,老年人骨小梁的异质性很强,并建议直接在临床或科学领域进行测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号