首页> 外文期刊>Otology and neurotology: official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology >Detection of bacterial biofilm on cochlear implants removed because of device failure, without evidence of infection.
【24h】

Detection of bacterial biofilm on cochlear implants removed because of device failure, without evidence of infection.

机译:耳蜗植入物上细菌生物膜的检测由于设备故障而被移除,没有感染迹象。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: To investigate the formation of bacterial biofilms on the surface of the electrode array of cochlear implants (CI) explanted because of device failure, without evidence of infection, by use of scanning electron microscopy (SEM) and confocal laser scanning microscopy (CLSM). STUDY DESIGN: Prospective study. SETTING: Patients from 2 tertiary-care referral centers. PATIENTS AND METHODS: CIs were explanted from 9 patients because of device failure. Specimens were immediately snap-frozen in cold isopenthane, stored at -80 degrees C and examined with SEM and CLSM by 3 investigators. MAIN OUTCOME MEASURE: Presence of bacterial biofilm ascertained by SEM and CSLM. RESULTS: One specimen showed the formation of a bacterial biofilm on the middle ear part of the electrode array. No biofilm formation was found in the inner-ear part of electrode arrays. In the middle-ear part of the electrode array, a cylindrical cover of human muscular tissue was seen plugging the cochleostomy. CONCLUSION: This is the first study demonstrating that bacterial biofilms may exist on the surface of the electrode array of CIs explanted because of device failure but not infection. We found 1 case of biofilm formation in 9 explanted CIs. Further studies with larger series of CIs are required to investigate biofilm formation on the surface of CI electrode arrays to address both the pathophysiology of bacterial biofilms and prevention of device-related infections in CI patients.
机译:目的:通过扫描电子显微镜(SEM)和共聚焦激光扫描显微镜(CLSM),研究由于装置故障而被植入的人工耳蜗植入物(CI)电极阵列表面上形成的细菌生物膜,而没有感染迹象。 。研究设计:前瞻性研究。地点:来自两个三级转诊中心的患者。患者与方法:由于器械故障,从9例患者中移出了CI。将样品立即在冷的异戊烷中速冻,保存在-80摄氏度下,并由3位研究人员用SEM和CLSM进行检查。主要观察指标:通过SEM和CSLM确定存在细菌生物膜。结果:一个标本显示在电极阵列的中耳部分形成细菌生物膜。在电极阵列的内耳部分未发现生物膜形成。在电极阵列的中耳部分,看到人肌肉组织的圆柱形覆盖物堵塞了耳蜗切开术。结论:这是第一项研究,证明被植入的CIs电极阵列表面可能存在细菌生物膜,这是由于装置故障而非感染引起的。我们在9个移出的CI中发现1例生物膜形成病例。需要对更大数量的CI进行进一步研究,以研究CI电极阵列表面上的生物膜形成,以解决CI患者中细菌生物膜的病理生理学和预防与设备相关的感染。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号