摘要

A special parallel plate ionization chamber, inserted in a slab phantom for the personal dose equivalent Hp(10) determination, was developed and characterized in this work. This ionization chamber has collecting electrodes and window made of graphite, and the walls and phantom made of PMMA. The tests comprise experimental evaluation following international standards and Monte Carlo simulations, employing the PENELOPE code to evaluate the design of this new dosimeter. The experimental tests were conducted employing the radioprotection level quality N-60 established at the IPEN, and all results were within the recommended standards. (C) 2015 Elsevier Ltd. All rights reserved.
机译:一个特殊的平行板电离室,插入到平板幻影中用于确定个人剂量当量Hp(10),并在这项工作中进行了表征。该电离室具有由石墨制成的收集电极和窗口,以及由PMMA制成的壁和模型。测试包括遵循国际标准和蒙特卡洛模拟的实验评估,并采用PENELOPE代码评估这种新型剂量计的设计。使用IPEN制定的防辐射等级N-60进行了实验测试,所有结果均在推荐标准之内。 (C)2015 Elsevier Ltd.保留所有权利。