首页> 外文期刊>Radiation measurements >Assessment of glandular dose and image quality in mammography using computerised radiography employing a polymethylmetacrilate breast simulator (Conference Paper)
【24h】

Assessment of glandular dose and image quality in mammography using computerised radiography employing a polymethylmetacrilate breast simulator (Conference Paper)

机译:乳腺X线摄影术中的腺剂量和图像质量评估,采用计算机放射成像技术,并使用聚甲基丙烯酸甲酯乳房模拟器(会议论文)

获取原文
获取原文并翻译 | 示例

摘要

In Brazil there are around 600 mammography equipment with CR system. Taking into account the quality of image is essential to evaluate the mean glandular dose so you can optimize the radiation protection of patients evaluated with this type of system. Therefore, this study aimed to determine the mean glandular dose and quality of image for the CR system of laboratory of radioprotection applied to mammography of the Centre of Development of Nuclear Technology (CDTN). For this, we evaluated the linearity of the detector's response, contrast to noise ratio and signal to noise ratio, which, according to European protocol, showed results within acceptable limits. Next, evaluated the quality of image with the CDMAM Phantom and mean glandular dose to the detector Fluke Biomedical TNT 12000WD, where they presented, respectively, within the results expected by the manufacturer and the limits of acceptable and desirable by the European protocol. then, the CR system of CDTN is optimized.
机译:在巴西,大约有600台带有CR系统的乳腺摄影设备。考虑图像质量对于评估平均腺体剂量至关重要,因此您可以优化使用这种类型系统评估的患者的放射防护。因此,本研究旨在确定应用于核技术发展中心(CDTN)的乳腺摄影的放射防护实验室CR系统的平均腺体剂量和图像质量。为此,我们评估了探测器响应的线性,与噪声之比和信噪比之比,根据欧洲协议,这些结果显示结果在可接受的范围内。接下来,使用CDMAM Phantom和Fluke Biomedical TNT 12000WD检测器评估了图像质量和平均腺体剂量,在制造商预期的结果范围内以及在欧洲协议可接受和期望的范围内,它们分别在此处提供。然后,对CDTN的CR系统进行了优化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号