首页> 外文期刊>IEEE Transactions on Instrumentation and Measurement >Quantum Calibration System for Digital Voltmeters at Voltages from 10 nV to 1 kV
【24h】

Quantum Calibration System for Digital Voltmeters at Voltages from 10 nV to 1 kV

机译:10 nV 至 1 kV 电压下数字电压表的量子校准系统

获取原文
获取原文并翻译 | 示例

摘要

We have developed a programmable Josephson voltage standard that can generate voltages up to 20 V with a resolution of 10 nV. The standard has two superconductor-normal metal-superconductor junction arrays connected in series and driven by two radio frequency oscillators. The standard has been integrated into a digital voltmeter calibration system for voltages from 10 nV up to 1000 V. The expanded uncertainty (k = 2) of the system is approximately 10 nV for voltages up to 20 V and 0.9 μV/V in the voltage range from 20 to 1000 V.
机译:我们开发了一种可编程约瑟夫森电压标准,可以产生高达 20 V 的电压,分辨率为 10 nV。该标准有两个串联的超导体-普通金属-超导体结阵列,并由两个射频振荡器驱动。该标准已集成到数字电压表校准系统中,电压范围为 10 nV 至 1000 V。对于高达20 V的电压,系统的扩展不确定度(k = 2)约为10 nV,在20至1000 V的电压范围内为0.9 μV/V。