Magnetic field gradients in the MRI suite and their effects on aneurysm clips.

机译:MRI套件中的磁场梯度及其对动脉瘤夹的影响。

获取原文
获取原文并翻译 | 示例
       

摘要

We studied magnetic field intensity in the magnetic resonance imaging suite at our hospital and its possible effect on several different types of aneurysm clips, including one Heifetz 17-7 PH, six Heifetz Elgiloy, one Mayfield, six Perneczky, fifteen Sugita, one Sundt-Kees Variangle, four Variangle-McFadden and fifteen Yasargil clips. We carefully observed the clips for any translational or rotational movements along the path from the door towards the magnetic resonance imaging gantry. The magnetic field strength was 0.04 kiloGauss at the entrance of the room, with an acute increase of magnetic strength at 310 cm away from the entrance to the room, 90 cm to the entrance of the gantry. The magnetic strength continued to increase at a rate of 1.0-1.5 kiloGauss for every 20 cm up to the entrance to the gantry. No movement was observed in any of the clips at the entrance to the suite except for the Heifetz 17-7 PH clip, which showed small movement in the longitudinal plane of the clip. At the entrance to the gantry, the Heifetz 17-7 PH, Sundt-Kees Variangle, and Mayfield clips were aligned on the walls of the test container perpendicular to the magnetic bore. The, Heifetz Elgiloy, Perneczky, Sugita, Variangle-McFadden, and Yasargil clips showed no movement throughout the path of the stretcher or near the gantry.
机译:我们在我们医院的磁共振成像套件中研究了磁场强度及其对几种不同类型的动脉瘤夹的可能影响,包括一个Heifetz 17-7 PH,六个Heifetz Elgiloy,一个Mayfield,六个Perneczky,十五个Sugita,一个Sundt- Kees Variangle,四个Variangle-McFadden和15个Yasargil剪辑。我们仔细观察了夹子在从门到磁共振成像机架的路径上是否存在任何平移或旋转运动。在房间入口处的磁场强度为0.04千高斯,在距房间入口310厘米,距龙门架入口90厘米处的磁场强度急剧增加。直到龙门架入口处,每20厘米的磁场强度继续以1.0-1.5千高斯的速度增加。除Heifetz 17-7 PH夹子外,在套房入口处的任何夹子中均未观察到运动,该夹子在夹子的纵向平面上显示出很小的运动。在机架的入口处,将Heifetz 17-7 PH,Sundt-Kees Variangle和Mayfield的夹子垂直于磁孔对准测试容器的壁。 Heifetz Elgiloy,Perneczky,Sugita,Variangle-McFadden和Yasargil剪辑在整个担架路径或龙门架附近均未显示移动。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号