首页> 外文期刊>Microelectronics international: Journal of ISHM--Europe, the Microelectronics Society--Europe >An investigation of the factors influencing stability in continuously-powered screen-printed amperometric dissolved oxygen sensors
【24h】

An investigation of the factors influencing stability in continuously-powered screen-printed amperometric dissolved oxygen sensors

机译:影响连续供电丝网印刷安培溶氧传感器稳定性的因素研究

获取原文
获取原文并翻译 | 示例
           

摘要

The continuously powered response of dissolved oxygen sensors fabricated as screen printed devices has been investigated. A reduction in the sensitivity of these devices has been observed when they are continuously powered at potentials of 0.6V versus Ag/AgCl reference electrodes. Periodically recorded voltammograms have indicated that the reduction in sensitivity can be attributed to a build up of hydroxyl ions beneath the sensor membrane that inhibits oxygen reduction. It has been found that powering these devices at more negative potentials mitigates against the decrease in sensitivity but results in slowly increasing sensitivity. THis is attributed to an increase in the effective area of the working electrode resulting mainly from leakage through the waterproofing layer covering the electrode interconnects.
机译:已经研究了制造为丝网印刷设备的溶解氧传感器的连续供电响应。与Ag / AgCl参比电极相比,以0.6V的电位连续供电时,已观察到这些设备的灵敏度降低。定期记录的伏安图已表明,灵敏度的降低可归因于传感器膜下方羟基离子的积累,从而抑制了氧气的还原。已经发现,以更大的负电势为这些设备供电减轻了灵敏度的降低,但是导致了灵敏度的缓慢提高。这归因于工作电极的有效面积增加,这主要是由于通过覆盖电极互连的防水层的泄漏引起的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号