2D/3D image fusion for accurate target localization and evaluation of a mask based stereotactic system in fractionated stereotactic radiotherapy of cranial lesions.

机译:2D / 3D图像融合,用于精确的目标定位和颅面病变的分级立体定向放射治疗中基于面罩的立体定向系统的评估。

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of this study was to evaluate the accuracy of a two-dimensional (2D) to three-dimensional (3D) image-fusion-guided target localization system and a mask based stereotactic system for fractionated stereotactic radiotherapy (FSRT) of cranial lesions. A commercial x-ray image guidance system originally developed for extracranial radiosurgery was used for FSRT of cranial lesions. The localization accuracy was quantitatively evaluated with an anthropomorphic head phantom implanted with eight small radiopaque markers (BBs) in different locations. The accuracy and its clinical reliability were also qualitatively evaluated for a total of 127 fractions in 12 patients with both kV x-ray images and MV portal films. The image-guided system was then used as a standard to evaluate the overall uncertainty and reproducibility of the head mask based stereotactic system in these patients. The phantom study demonstrated that the maximal random error of the image-guided target localization was +/-0.6 mm in each direction in terms of the 95% confidence interval (CI). The systematic error varied with measurement methods. It was approximately 0.4 mm, mainly in the longitudinal direction, for the kV x-ray method. There was a 0.5 mm systematic difference, primarily in the lateral direction, between the kV x-ray and the MV portal methods. The patient study suggested that the accuracy of the image-guided system in patients was comparable to that in the phantom. The overall uncertainty of the mask system was +/-4 mm, and the reproducibility was +/-2.9 mm in terms of 95% CI. The study demonstrated that the image guidance system provides accurate and precise target positioning.
机译:这项研究的目的是评估二维(2D)到三维(3D)图像融合引导目标定位系统和基于掩码的立体定向系统用于颅骨病变的分级立体定向放射治疗(FSRT)的准确性。最初为颅外放射外科手术开发的商业X射线图像引导系统用于颅骨病变的FSRT。通过在不同位置植入了八个小的不透射线的标记物(BBs)的拟人化头部模型来定量评估定位精度。还对12例kV x射线图像和MV门静脉片的患者中的127个部分的准确性和临床可靠性进行了定性评估。然后将图像引导系统用作评估这些患者基于头罩的立体定向系统的总体不确定性和可重复性的标准。幻像研究表明,按照95%置信区间(CI),每个方向上图像引导目标定位的最大随机误差为+/- 0.6 mm。系统误差随测量方法而变化。对于kV x射线方法,主要在纵向方向上约为0.4mm。 kV x射线方法和MV门户方法之间的系统差异主要在横向上有0.5 mm。病人研究表明,病人体内影像引导系统的准确性与体模中的准确性相当。掩模系统的总体不确定度为+/- 4 mm,就95%CI而言,重现性为+/- 2.9 mm。研究表明,图像制导系统可提供准确而精确的目标定位。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号