Consistency of breast density measured from the same women in four different MR scanners

机译:在四台不同的MR扫描仪中从同一位妇女测量的乳房密度的一致性

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Purpose: To compare the breast volume (BV), fibroglandular tissue volume (FV), and percent density (PD) measured from breast MRI of the same women using four different MR scanners. Methods: The study was performed in 34 healthy Asian volunteers using two 1.5T (GE and Siemens) and two 3T (GE and Philips) MR scanners. The BV, FV, and PD were measured on nonfat-suppressed T1-weighted images using a comprehensive computer algorithm-based segmentation method. The scanner-to-scanner measurement difference, and the coefficient of variation (CV) among the four scanners were calculated. The measurement variation between two density morphological patterns presenting as the central type and the intermingled type was separately analyzed and compared. Results: All four scanners provided satisfactory image quality allowing for successful completion of the segmentation processes. The measured parameters between each pair of MR scanners were highly correlated, with R 2 ≥0.95 for BV, R 2 ≥0.99 for FV, and R 2 ≥ 0.97 for PD in all comparisons. The mean percent differences between each pair of scanners were 5.9-7.8 for BV, 5.3-6.5 for FV, 4.3-7.3 for PD; with the overall CV of 5.8 for BV, 4.8 for FV, and 4.9 for PD. The variation of FV was smaller in the central type than in the intermingled type (p 0.04). Conclusions: The results showed that the variation of FV and PD measured from four different MR scanners is around 5, suggesting the parameters measured using different scanners can be used for a combined analysis in a multicenter study.
机译:目的:比较使用四个不同的MR扫描仪对同一位妇女的乳房MRI测量的乳房体积(BV),纤维腺组织体积(FV)和密度百分比(PD)。方法:本研究在34名亚洲健康志愿者中使用两台1.5T(GE和Siemens)和两台3T(GE和Philips)MR扫描仪进行。使用基于计算机算法的综合分割方法,在非脂肪抑制的T1加权图像上测量BV,FV和PD。计算出扫描仪之间的测量差异以及四个扫描仪之间的变异系数(CV)。分别分析和比较了呈现为中心型和混合型的两种密度形态模式之间的测量变化。结果:所有四台扫描仪均提供令人满意的图像质量,可成功完成分割过程。在所有比较中,每对MR扫描仪之间的测量参数高度相关,BV的R 2≥0.95,FV的R 2≥0.99,PD的R 2≥0.97。每对扫描仪之间的平均百分比差异对于BV为5.9-7.8,对于FV为5.3-6.5,对于PD为4.3-7.3; BV的总CV为5.8,FV的总CV为4.8,PD的总CV为4.9。 FV的变化在中央型比混合型小(p 0.04)。结论:结果表明,从四个不同的MR扫描仪测得的FV和PD的变化约为5,这表明使用不同的扫描仪测得的参数可用于多中心研究的组合分析。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号