We have developed a liquid chromatography tandem mass spectrometry (LC-MS/MS) system capable of achieving better than 2% accuracy, routinely over a wide concentration range of 1-800 ngmL(-1). We demonstrate that the necessary high precision, high accuracy and rapid analysis can be achieved using LC-MS/MS technology. Automated nanoelectrospray ionisation tandem mass spectrometry (nanoESI-MS/MS) technology can be employed to eliminate the chromatographic step completely. In this paper, nanoESI-MS/MS is evaluated and compared directly with LC-MS/MS for the quantitative analysis of two-test analytes, amitriptyline (ATT) and 5-methoxytryptamine (5-MTT), in aqueous/organic mixture. Calibration curves were found to be linear over a wide concentration range of 1-800 ngmL(-1) for both analytes using LC-MS/MS. Using nanoESI-MS/MS ATT gave a linear response while 5-MTT gave a non-linear response using nanoESI-MS/MS over the same concentration range as in LC-MS/MS. Accuracy and precision values of quality control samples (QCs) at four concentration levels were analysed in replicates of six at each level using 5-MTT and ATT as test analytes for both techniques. The LC-MS/MS system was capable of achieving accuracy levels of 99.50-101.96% for ATT and 100.17-100.40% for 5-MTT. Accuracy levels using nanoESI-MS/MS were not comparable to LC-MS/MS, they ranged from 90.09-100.18% for ATT and 95.95-113.55% for 5-MTT. The precision values obtained for nanoESI-MS/MS were in good agreement with those obtained by LC-MS/MS.
机译:我们已经开发了一种液相色谱串联质谱分析(LC-MS / MS)系统,该系统能够在1-800 ngmL(-1)的宽浓度范围内常规获得优于2%的准确度。我们证明使用LC-MS / MS技术可以实现必要的高精度,高精度和快速分析。可以采用自动化的纳米电喷雾电离串联质谱(nanoESI-MS / MS)技术来完全消除色谱步骤。在本文中,对nanoESI-MS / MS进行了评估,并将其与LC-MS / MS直接进行了定量比较,以分析含水/有机混合物中的两次测试分析物阿米替林(ATT)和5-甲氧基色胺(5-MTT)。发现使用LC-MS / MS两种分析物在1-800 ngmL(-1)的宽浓度范围内的校准曲线都是线性的。使用nanoESI-MS / MS,在与LC-MS / MS相同的浓度范围内,使用nanoESI-MS / MS ATT可以得到线性响应,而使用5-MTT可以给出非线性响应。使用5-MTT和ATT作为两种技术的测试分析物,对四个浓度水平的质量控制样品(QC)的准确性和精密度值进行了分析,每个浓度重复六次。 LC-MS / MS系统的ATT精度水平为99.50-101.96%,5-MTT的精度为100.17-100.40%。使用nanoESI-MS / MS的准确性水平无法与LC-MS / MS相提并论,ATT的准确度范围为90.09-100.18%,5-MTT的准确度范围为95.95-113.55%。 nanoESI-MS / MS获得的精度值与LC-MS / MS获得的精度值非常一致。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号