Reproducibility of real-time shear wave elastography in the evaluation of liver elasticity

机译:实时剪切波弹性成像在肝弹性评估中的重现性

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Objective: To evaluate the reproducibility of real-time shear wave elastography in assessing liver elasticity in healthy volunteers. Methods: Forty-two volunteers were studied in day 1. Shear wave elastography studies were performed by using the ultrasound system Aixplorer? (SuperSonic Imagine S.A., Aix-en-Provence, France) with a convex broadband probe. Measurements were carried by two operators, an expert (operator 1) and a novice (operator 2). Examinations were performed on the right lobe of the liver. Each operator performed 10 consecutive measurements in each volunteer. In a subset of volunteers (n = 18) measurements were performed twice on two different days (day 1 and day 2). Intraobserver and interobserver agreement were assessed by intraclass correlation coefficient. Results: Intraobserver agreement between measurements performed in the same subject in the same day (day 1 or day 2) showed intraclass correlation coefficient values of 0.95 (95% confidence interval, 0.93-0.98) and 0.93 (95% confidence interval, 0.90-0.96) for operator 1 and operator 2, respectively. Intraobserver agreement between measurements performed in the same subject in different days showed intraclass correlation coefficient values of 0.84 (95% confidence interval, 0.69-0.98) and 0.65 (95% confidence interval, 0.39-0.91) for operator 1 and operator 2, respectively. Interobserver agreement was 0.88 (95% confidence interval, 0.82-0.94). Conclusions: The results of this study show that shear wave elastography is a reliable and reproducible noninvasive method for the assessment of liver elasticity. Expert operator had higher reproducibility of measurements over time than novice operator.
机译:目的:评估实时剪切波弹性成像技术在评估健康志愿者肝脏弹性方面的可重复性。方法:在第1天研究了42名志愿者。使用超声系统Aixplorer?进行了剪切波弹性成像研究。 (SuperSonic Imagine S.A.,法国普罗旺斯),带有凸形宽带探头。测量由两名操作员进行,专家(操作员1)和新手(操作员2)。检查在肝右叶进行。每个操作员在每个志愿者中进行10次连续测量。在一部分志愿者(n = 18)中,在两个不同的天(第1天和第2天)进行了两次测量。观察者内部和观察者之间的一致性通过类内相关系数进行评估。结果:在同一天(第1天或第2天)对同一受试者进行的测量之间的观察员内部一致性表明,组内相关系数值分别为0.95(95%置信区间,0.93-0.98)和0.93(95%置信区间,0.90-0.96) )分别用于运算符1和运算符2。观察者之间的同一天在不同日期进行的测量之间的观察者内一致性表明,操作员1和操作员2的组内相关系数值分别为0.84(95%置信区间,0.69-0.98)和0.65(95%置信区间,0.39-0.91)。观察者之间的一致性为0.88(95%置信区间0.82-0.94)。结论:这项研究的结果表明,剪切波弹性成像是一种可靠且可重现的无创方法,用于评估肝脏弹性。与新手相比,专业操作员随着时间的推移具有更高的测量重现性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号