Lifting disabled patients onto the densitometer with a ceiling lift: Effect of the sling on measurement of BMD

机译:用天花板升降机将残疾患者抬到密度计上:吊索对BMD测量的影响

获取原文
获取原文并翻译 | 示例
           

摘要

Objective: Lifting disabled patients onto a densitometer manually is dangerous for both the patient and the densitometry staff; using a ceiling lift is the preferred method of transfer. This system requires the use of a sling underneath the patient. Unless extra time is taken for its removal, the sling remains underneath the patient as bone mineral density (BMD) is measured. The aim of this study was to determine whether leaving this sling in place during scan acquisition affects the BMD measurement.Methods: Measurements were taken of a spine phantom 30 times by itself, 30 times with a standard sling underneath the spine phantom, and 16 times with a disposable sling underneath the spine phantom.Results: We found that mean BMD was significantly different versus the phantom alone when a sling was used, due to differences in area, bone mineral content, or both. The disposable sling affected the mean BMD to a much greater extent than did the standard sling (+1.9% vs. -0.41%; P for the difference between slings <.001). The standard sling did not increase the variance in the BMD measurement compared with the spine phantom alone, whereas the disposable sling did increase the variance in the BMD measurements.Conclusion: Commercially available ceiling-lift slings affect BMD measurements of spine phantoms. This effect is expected to persist when BMD is measured in patients and suggests that when lifting a patient onto the densitometer using these slings, it is best to take the time to remove the sling from under the patient after transfer and before scanning.
机译:目的:手动将残疾患者抬到密度计上对患者和密度测定人员都是危险的;首选吊车是转移的方法。该系统需要在患者下方使用吊索。除非花费额外的时间来去除它,否则在测量骨矿物质密度(BMD)时,将悬带留在患者下方。这项研究的目的是确定在扫描采集过程中将悬带放置在适当的位置是否会影响BMD测量。方法:对脊柱体模进行30次自身测量,对标准体模在脊柱体模下面进行30次测量,对16次进行测量结果:我们发现,由于面积,骨矿物质含量或两者之间的差异,使用吊带时,平均BMD与仅使用幻像时的BMD显着不同。一次性吊索对平均BMD的影响程度比标准吊索大得多(+ 1.9%对-0.41%; P表示吊索之间的差异<.001)。与单独的脊柱体模相比,标准悬带不会增加BMD测量值的差异,而一次性悬带确实会增加BMD测量值的变化。结论:市售的吊顶吊索会影响脊柱体模的BMD测量值。当在患者中测量BMD时,预期这种效果会持续存在,并建议在使用这些吊索将患者抬到光密度计上时,最好花些时间在转移后和扫描之前从患者下方取出吊索。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号