Do ultrasound measurements reflect bone microarchitecture rather than bone mass?: An in vitro study of the rat femur with the use of ultrasound, densitometry, and histomorphometry.

机译:超声测量能反映出骨骼的微结构而不是骨骼质量吗?:使用超声,光密度法和组织形态学对大鼠股骨的体外研究。

获取原文
获取原文并翻译 | 示例

摘要

RATIONALE AND OBJECTIVES: In an experimental study in 40 rat femurs, the authors correlated the amplitude-dependent speed of bone ultrasound (Ad-SOS) with the bone mineral content and density and with the bone trabecular connectivity: trabecular bone volume, trabecular number, trabecular thickness, and trabecular separation to evaluate and compare the usefulness of the Ad-SOS to determine bone quantity and/or quality. METHODS: Bone mineral content and density were determined with dual-energy x-ray absorptiometry. Trabecular connectivity was determined with histomorphometric techniques. RESULTS: There was a strong correlation between the Ad-SOS and the other parameters studied, with a particularly high positive correlation with trabecular bone volume and trabecular thickness, and an inverse correlation with trabecular separation. The correlation was weaker with the bone mineral content and bone mineral density and with the trabecular number. For the trabecular separation, the correlation was significant in all cases, but it was negative. CONCLUSIONS: Bone ultrasound, in this case Ad-SOS, defines the quality of the bone in terms of trabecular architecture rather than bone density; however, this conclusion is valid only for the rat femur model that the authors used.
机译:理由和目的:在一项针对40只大鼠股骨的实验研究中,作者将幅度依赖的骨超声(Ad-SOS)速度与骨矿物质含量和密度以及骨小梁连接性相关:小梁骨体积,小梁数目,骨小梁厚度和骨小梁分离,以评估和比较Ad-SOS对确定骨量和/或质量的有用性。方法:采用双能X线吸收法测定骨矿物质含量和密度。用组织形态计量学技术确定小梁连接性。结果:Ad-SOS与研究的其他参数之间具有很强的相关性,与骨小梁的骨体积和骨小梁的厚度呈正相关,与骨小梁的分离呈负相关。骨矿物质含量,骨矿物质密度和骨小梁数目之间的相关性较弱。对于小梁分离,相关性在所有情况下均显着,但为负相关。结论:骨超声,在这种情况下为Ad-SOS,根据小梁结构而不是骨密度来定义骨的质量。但是,该结论仅对作者使用的大鼠股骨模型有效。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号