Positioning accuracy for lung stereotactic body radiotherapy patients determined by on-treatment cone-beam CT imaging

机译:通过治疗中的锥形束CT成像确定肺部立体定向放疗患者的定位精度

获取原文
获取原文并翻译 | 示例
       

摘要

Stereotactic body radiotherapy for early stage non-small cell lung cancer is an emerging treatment option in the UK. Since relatively few high-dose ablative fractions are delivered to a small target volume, the consequences of a geometric miss are potentially severe. This paper presents the results of treatment delivery set-up data collected using Elekta Synergy (Elekta, Crawley, UK) cone-beam CT imaging for 17 patients immobilised using the Bodyfix system (Medical Intelligence, Schwabmuenchen, Germany). Images were acquired on the linear accelerator at initial patient treatment set-up, following any position correction adjustments, and posttreatment. These were matched to the localisation CT scan using the Elekta XVI software. In total, 71 fractions were analysed for patient set-up errors. The mean vector error at initial set-up was calculated as 5.3±2.7mm, which was significantly reduced to 1.4±0.7mm following image guided correction. Post-treatment the corresponding value was 2.1±1.2mm. The use of the Bodyfix abdominal compression plate on 5 patients to reduce the range of tumour excursion during respiration produced mean longitudinal set-up corrections of -4.4±4.5mm compared with -0.7±2.6mm without compression for the remaining 12 patients. The use of abdominal compression led to a greater variation in set-up errors and a shift in the mean value.
机译:立体定向放射疗法用于早期非小细胞肺癌是英国新兴的治疗选择。由于将相对较少的高剂量消融级分输送到较小的目标体积,因此几何错位的后果可能很严重。本文介绍了使用Elekta Synergy(英国Elekta,Crawley,英国)锥束CT成像收集的治疗交付设置数据的结果,这些数据通过Bodyfix系统(医疗智能,Schwabmuenchen,德国)固定了17例患者。在进行任何位置校正调整和后处理之后,在初始患者治疗设置时,使用线性加速器获取图像。使用Elekta XVI软件将这些与定位CT扫描相匹配。总共分析了71个馏分的患者设置错误。初始设置时的平均矢量误差经计算为5.3±2.7mm,在图像引导校正后已显着降低至1.4±0.7mm。处理后的相应值为2.1±1.2mm。在5名患者上使用Bodyfix腹部加压板来减少呼吸过程中肿瘤偏移的范围,使其余12名患者在不加压的情况下,平均纵向安装矫正为-4.4±4.5mm,而没有加压的为-0.7±2.6mm。腹部按压的使用导致设置误差的更大变化和平均值的变化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号