Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube.

机译:气管导管袖口漏液:评估新的Microcuff气管导管。

获取原文
获取原文并翻译 | 示例

摘要

OBJECTIVE. This study compared the recently introduced Microcuff endotracheal tube HVLP ICU featuring an ultrathin (7- micro m) polyurethane cuff membrane with endotracheal tubes from different manufacturers regarding fluid leakage past the tube cuff. DESIGN. In vitro setup. MEASUREMENTS AND RESULTS. The following endotracheal tubes (ID 7.5 mm) were compared: Mallinckrodt HiLo, Microcuff HVLP ICU, Portex Profile Soft Seal, Rusch Super Safety Clear, and Sheridan CF. A vertical PVC trachea model (ID 20 mm) was intubated, and cuffs were inflated to 10, 15, 20, 25, 30, and 60 cmH(2)O. Colored water (5 ml) was added to the top of the cuff. The amount of leaked fluid past the tube cuff within 5, 10, and 60 min was recorded. Experiments were performed four times using two examples of each tube brand. Fluid leakage past tube cuffs occurred in all conventional endotracheal tubes at cuff pressures from 10 to 60 cmH(2)O. In the Microcuff tube cuff pressure fluid leakage was observed within 10 min only at 10 cmH(2)O. Results with the Microcuff tube were significantly better than all other tube brands at cuff pressures of 10-30 cmH(2)O. CONCLUSIONS. Within the acceptable upper limit for tracheal cuff pressure (25-30 cmH(2)O) the Microcuff endotracheal tube was the only one of the tested tubes to prevent fluid leakage in our in vitro setup. In vivo studies are required to confirm these findings.
机译:目的。这项研究比较了最近推出的具有超薄(7微米)聚氨酯袖带膜的Microcuff气管插管HVLP ICU和来自不同制造商的气管插管在流体通过管套时的泄漏情况。设计。体外设置。测量和结果。比较了以下气管插管(内径7.5 mm):Mallinckrodt HiLo,Microcuff HVLP ICU,Portex Profile软密封,Rusch Super Safety Clear和SheridanCF。插入垂直的PVC气管模型(ID 20 mm),并将袖带充气至10、15、20、25、30和60 cmH(2)O。将有色水(5 ml)添加到袖带的顶部。记录在5、10和60分钟内通过管套泄漏的液体量。使用每个试管品牌的两个示例进行了四次实验。在10到60 cmH(2)O的袖带压力下,所有常规气管插管中都会发生通过袖带漏液的情况。在Microcuff管袖带压力中,仅在10 cmH(2)O下10分钟内观察到流体泄漏。在10-30 cmH(2)O的袖带压力下,使用Microcuff管的结果显着优于所有其他管品牌。结论。在气管袖带压力的可接受上限(25-30 cmH(2)O)内,Microcuff气管插管是唯一一种在我们的体外设置中可以防止流体泄漏的被测管。需要进行体内研究以证实这些发现。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号