Determination of plasma free carnitine by electrospray tandem mass spectrometry.

机译:电喷雾串联质谱法测定血浆游离肉碱。

获取原文
获取原文并翻译 | 示例
           

摘要

We have evaluated a tandem mass spectrometry method for determining free carnitine concentrations in plasma and have compared its performance with that of an existing radioenzymatic assay. In this method, plasma was mixed with an internal standard, carnitine-d3 (600 nmol/L), and butylated before analysis on a Quattro II tandem mass spectrometer. The detection limit of the tandem mass spectrometric (MS) method was 4 micromol/L and carryover between samples was 2.2%. Precision of the method was 1.6-5.4% between injections and 4.0-5.6% between samples. Between-batch precision was 10-13%. The method was linear up to a free carnitine concentration of 300 micromol/L and good agreement was found with the existing radioenzymatic assay. We conclude that the tandem MS method is a precise and robust method for the determination of free carnitine concentrations in plasma that overcomes the disadvantages of the radiochemical method.
机译:我们已经评估了串联质谱法测定血浆中游离肉碱的浓度,并将其性能与现有的放射性酶法进行了比较。在这种方法中,将血浆与内标肉碱-d3(600 nmol / L)混合,并进行丁基化,然后在Quattro II串联质谱仪上进行分析。串联质谱(MS)方法的检出限为4 micromol / L,样品之间的残留率为2.2%。该方法的精密度在进样之间为1.6-5.4%,样品之间为4.0-5.6%。批间精度为10-13%。该方法在游离肉碱浓度高达300 micromol / L时是线性的,与现有的放射性酶法检测结果吻合良好。我们得出结论,串联质谱法是一种精确而可靠的测定血浆中游离肉碱浓度的方法,克服了放射化学方法的缺点。