Comparative study on auscultatory and oscillometric methods of ambulatory blood pressure measurements in adult patients.

机译:成人患者动态血压的听诊和示波法的比较研究。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

BACKGROUND: Ambulatory blood pressure (BP) monitors measure BP by oscillometric (Osc) or auscultatory (Aus) methods. In this study, we aimed to compare the values obtained by BP measurement methods in the ambulatory blood pressure monitoring (ABPM) device that can take measurements using both the methods, factors influencing these values, and their relevancy in obtaining valid measurement percentages. METHODS: The study was conducted on 212 patients. BP values of these patients were measured with the Nissei DS-250 (Nihon Seimitsu Sokki Co. Ltd, Gunma, Japan) monitor capable of recording BP simultaneously by the Aus and the Osc methods. The cases, for which no records were obtained by any one of the two methods, were considered to have missing records. The impacts of age, sex, body mass index, and arm circumference on the valid BP measurement values were examined. RESULTS: The valid measurement percentage without any missing records obtained by the Osc method was 87.3 and 16% by the Aus method. The bodymass index, age, and arm circumference did not affect this percentage. The 24-h BP measurement values were higher by the Osc method. The differences between both methods in systolic blood pressure and diastolic blood pressure were 3.8+/-4.6 and 2.3+/-2.7 mmHg, respectively. Although the Aus method yielded higher standard deviation values than the Osc method for systolic blood pressure, standard deviation values obtained by the Osc method were higher than those obtained by the Aus method for diastolic blood pressure. CONCLUSION: The percentage of obtaining valid measurements by the Osc method used in the Nissei DS-250 model ABPM device examined in this study is higher when compared with the Aus method. Therefore, while examining ABPM results, the Osc method may be preferable over the Aus method.
机译:背景:动态血压(BP)监视器通过示波法(Osc)或听诊(Aus)方法测量血压。在本研究中,我们旨在比较动态血压监测(ABPM)设备中通过BP测量方法获得的值,该设备可以使用两种方法,影响这些值的因素及其在获取有效测量百分比中的相关性进行测量。方法:该研究针对212名患者进行。用能够同时通过Aus和Osc方法记录血压的Nissei DS-250(日本群光株式会社Nihon Seimitsu Sokki Co.Ltd)测量这些患者的BP值。通过两种方法中的任何一种均未获得任何记录的案件被视为缺少记录。检查了年龄,性别,体重指数和手臂围度对有效血压测量值的影响。结果:通过Osc方法获得的有效测量百分比无丢失记录的有效百分比为87.3,通过Aus方法获得的有效度量百分比为16%。身体质量指数,年龄和手臂围度均不影响该百分比。通过Osc方法测得的24小时BP测量值较高。两种方法的收缩压和舒张压之间的差异分别为3.8 +/- 4.6和2.3 +/- 2.7 mmHg。尽管对于收缩压,Aus方法产生的标准偏差值要高于Osc方法,但通过Osc方法获得的标准偏差值要高于通过Aus方法获得的舒张血压标准偏差值。结论:本研究研究的Nissei DS-250型ABPM设备中使用Osc方法获得有效测量值的百分比与Aus方法相比更高。因此,在检查ABPM结果时,Osc方法可能优于Aus方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号