首页> 外文期刊>Blood Pressure Monitoring >Effect of the shapes of the oscillometric pulse amplitude envelopes and their characteristic ratios on the differences between auscultatory and oscillometric blood pressure measurements.
【24h】

Effect of the shapes of the oscillometric pulse amplitude envelopes and their characteristic ratios on the differences between auscultatory and oscillometric blood pressure measurements.

机译:示波脉冲幅度包络的形状及其特征比对听诊和示波血压测量值之间差异的影响。

获取原文
获取原文并翻译 | 示例
       

摘要

INTRODUCTION: Oscillometric noninvasive blood pressure (NIBP) devices determine pressure by analysing the oscillometric waveform using empirical algorithms. Many algorithms analyse the waveform by calculating the systolic and diastolic characteristic ratios, which are the amplitudes of the oscillometric pulses in the cuff at, respectively, the systolic and diastolic pressures, divided by the peak pulse amplitude. A database of oscillometric waveforms was used to study the influences of the characteristic ratios on the differences between auscultatory and oscillometric measurements. METHODS: Two hundred and forty-three oscillometric waveforms and simultaneous auscultatory blood pressures were recorded from 124 patients at cuff deflation rates of 2-3 mmHg/s. A simulator regenerated the waveforms, which were presented to two NIBP devices, the Omron HEM-907 [OMRON Europe B.V. (OMCE), Hoofddorp, The Netherlands] and the GE ProCare 400 (GE Healthcare, Tampa, Florida, USA). For each waveform, the paired systolic and paired diastolic pressure differences between device measurements and auscultatory reference pressures were calculated. The systolic and diastolic characteristic ratios, corresponding to the reference auscultatory pressures of each oscillometric waveform stored in the simulator, were calculated. The paired differences between NIBP measured and auscultatory reference pressures were compared with the characteristic ratios. RESULTS: The mean and standard deviations of the systolic and diastolic characteristic ratios were 0.49 (0.11) and 0.72 (0.12), respectively. The systolic pressures recorded by both devices were lower (negative paired pressure difference) than the corresponding auscultatory pressures at low systolic characteristic ratios, but higher than the corresponding auscultatory pressures at high systolic pressures. Conversely, the differences between the paired diastolic pressure differences were higher at low diastolic characteristic ratios, compared with those at high diastolic characteristic ratios. The paired systolic pressure differences were within +/-5 mmHg for those waveforms with systolic characteristic ratios between 0.4 and 0.7 for the Omron and between 0.3 and 0.5 for the ProCare. The paired diastolic pressure differences were within +/-5 mmHg for those waveforms with diastolic characteristic ratios between 0.4 and 0.6 for the Omron and between 0.5 and 0.8 for the ProCare. DISCUSSION AND CONCLUSION: The systolic and diastolic paired oscillometric-auscultatory pressure differences varied with their corresponding characteristic ratios. Good agreement (within 5 mmHg) between the oscillometric and auscultatory pressures occurred for oscillometric pulse amplitude envelopes with specific ranges of characteristic ratios, but the ranges were different for the two devices. Further work is required to classify the different envelope shapes, comparing them with patient conditions, to determine if a clearer understanding of the different waveform shapes would improve the accuracy of oscillometric measurements.
机译:简介:示波无创血压(NIBP)设备通过使用经验算法分析示波波形来确定压力。许多算法通过计算收缩压和舒张压的特征比来分析波形,这是分别在收缩压和舒张压下的袖带中示波脉冲的振幅除以峰值脉冲振幅。使用示波波形数据库来研究特征比对听诊和示波测量之间的差异的影响。方法:记录124例患者的243个示波波形和同时听诊血压,袖带放气率为2-3 mmHg / s。模拟器重新生成了波形,并提供给两个NIBP设备:Omron HEM-907 [Omron Europe B.V.(OMCE),荷兰霍夫多普]和GE ProCare 400(GE Healthcare,美国佛罗里达州坦帕市)。对于每个波形,计算出设备测量值与听诊参考压力之间的成对的收缩压和成对的舒张压差。计算出相应于模拟器中存储的每个示波波形的参考听诊压力的收缩和舒张特性比。将NIBP测得的和听诊参考压力之间的成对差异与特征比率进行比较。结果:收缩和舒张特征比的平均值和标准差分别为0.49(0.11)和0.72(0.12)。在低收缩特性比率下,两个装置记录的收缩压均低于相应的听诊压力(负配对压力差),但高于高收缩压下相应的听诊压力。相反,与低舒张压特性比相比,成对的舒张压差之间的差在低舒张特性比时更高。对于这些波形,成对的收缩压差在+/- 5 mmHg范围内,对于Omron而言收缩特性比在0.4和0.7之间,对于ProCare而言,收缩特性比在0.3和0.5之间。对于这些波形,成对的舒张压差在+/- 5 mmHg范围内,其中Omron的舒张特性比在0.4和0.6之间,ProCare的舒张特性比在0.5和0.8之间。讨论与结论:收缩压与舒张压成对的示波-听诊压力差随其相应的特征比而变化。对于具有特定比例特征比的示波脉冲幅度包络,示波压力和听诊压力之间达成了良好的一致性(在5 mmHg之内),但是两种设备的范围不同。需要进一步的工作来对不同的包络形状进行分类,并将其与患者的状况进行比较,以确定对不同波形形状的更清晰的了解是否会提高示波法测量的准确性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号