Systematic errors in isothermal titration calorimetry: Concentrations and baselines

机译:等温滴定量热法的系统误差:浓度和基线

获取原文
获取原文并翻译 | 示例
           

摘要

In the study of 1:1 binding by isothermal titration calorimetry, reagent concentration errors are fully absorbed in the data analysis, giving incorrect values for the key parameters - K, ΔH, and n - with no effect on the least-squares statistics. Reanalysis of results from an interlaboratory study of a selected biochemical process demonstrates that concentration errors are likely responsible for most of the overall statistical error in these parameters. The concentration errors are approximately 10%, greatly exceeding expected levels. Furthermore, examination of selected data sets reveals a surprising sensitivity to the baseline, suggesting a need for great care in treating dilution heats.
机译:在通过等温滴定量热法进行1:1结合的研究中,试剂浓度误差在数据分析中被完全吸收,给出了关键参数K,ΔH和n的不正确值,而对最小二乘统计量没有影响。对所选生化过程的实验室间研究结果的重新分析表明,浓度误差很可能是这些参数中大部分总体统计误差的原因。浓度误差约为10%,大大超出了预期水平。此外,对选定数据集的检查显示出对基线的惊人的敏感性,这表明在处理稀释热时需要格外小心。