首页> 外文期刊>Analytical and bioanalytical chemistry >Instrumental modification intended to save time, and volumes of sample and reagent solutions, in the atomic fluorescence spectrometric determination of mercury
【24h】

Instrumental modification intended to save time, and volumes of sample and reagent solutions, in the atomic fluorescence spectrometric determination of mercury

机译:仪器改造旨在节省时间,并节省原子荧光光谱法测定汞的样品和试剂溶液的体积

获取原文
获取原文并翻译 | 示例
           

摘要

Use of small membrane pumps, instead of peristaltic pumps, to introduce sample and reagent solutions into the spectrometer has several advantages in atomic fluorescence spectrometric determination of mercury. This simple modification results in a substantial saving in the time required for the measurements and so 90% of reagent solution volumes and 95% of sample solution volumes are saved, with a consequent decrease in the volume of waste generated. The sampling frequency is almost tripled, with no deterioration in sensitivity, which is similar to that obtained by use of peristaltic pumps. The relative standard deviation for ten consecutive measurements of a 1 mu g L-1 mercury solution was approximately 2%.
机译:在原子荧光光谱法测定汞中,使用小型膜泵代替蠕动泵将样品和试剂溶液引入光谱仪具有多个优点。这种简单的修改大大节省了测量所需的时间,因此节省了90%的试剂溶液体积和95%的样品溶液体积,从而减少了产生的废物量。采样频率几乎增加了两倍,灵敏度没有降低,这与使用蠕动泵获得的频率相似。 1μgL-1汞溶液的十次连续测量的相对标准偏差约为2%。