首页> 外文期刊>Analytical chemistry >Real-time determination of aromatics in automobile exhaust by single photon ionisation ion trap mass spectrometry
【24h】

Real-time determination of aromatics in automobile exhaust by single photon ionisation ion trap mass spectrometry

机译:单光子电离离子阱质谱法实时测定汽车尾气中的芳烃

获取原文
获取原文并翻译 | 示例
           

摘要

Instrumentation was constructed for the determination of oxygenates and aromatics by single-photon ionization ion trap mass spectrometry (SPI-ITMS) with a direct-sampling interface for application to real-time automotive exhaust measurements. Vacuum ultraviolet light at 118 mn was produced by focusing 355-nm pulses fi om a Nd:YAG laser into a gas cell containing xenon and passing them through ports in the ring electrode of a modified commercial ion trap mass spectrometer. The direct-sampling system consisted of a heated capillary interface with a pulsed valve. Good ionization efficiency was observed for analytes with ionization potentials (IPs) below 10.1 eV, with most molecules yielding molecular ions as the most abundant ionization product. SPI provides significant selectivity against interferences with IPs greater than 10.5 eV, Tandem mass spectrometry was employed to distinguish between analytes that produced isobaric base peaks via SPI. Detection limits between 0.03 and 0.2 ppm were obtained for seven compounds by SPI, with precision of 50% at 1 ppm, a time resolution of 5 s, and a linear dynamic range of 2-3 orders of magnitude. Accuracy was verified by the simultaneous determination of six compounds at similar to 1 ppm in a gas mixture standard. Real-time analysis of automobile exhaust by SPI-ITMS demonstrated the potential of the system for emissions measurements. [References: 34]
机译:构造了用于通过具有直接采样接口的单光子电离离子阱质谱(SPI-ITMS)测定含氧化合物和芳族化合物的仪器,用于实时汽车尾气测量。通过将355 nm的脉冲从Nd:YAG激光聚焦到包含氙气的气室中,并使它们穿过改良型工业离子阱质谱仪的环形电极中的端口,产生118 mn的真空紫外光。直接采样系统由加热的毛细管接口和脉冲阀组成。对于电离电势(IPs)低于10.1 eV的分析物,观察到良好的电离效率,大多数分子产生的分子离子是最丰富的电离产物。 SPI提供了对大于10.5 eV的IP干扰的显着选择性,采用串联质谱法来区分通过SPI产生等压碱基峰的分析物。通过SPI获得的7种化合物的检出限在0.03至0.2 ppm之间,在1 ppm处的精密度为50%,时间分辨率为5 s,线性动态范围为2-3个数量级。通过同时测定六种化合物(在混合气体标准溶液中以接近1 ppm的浓度)来验证准确性。通过SPI-ITMS对汽车尾气进行实时分析,证明了该系统在排放测量中的潜力。 [参考:34]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号