In this paper, a method is described for the chronoamperometric determination of Ce(IV) in acidic aqueous solutions in order to study the kinetics and the mechanism of polymerization reactions with Ce(IV) as the initiator. The method is based on the FIA principle, in which small samples are injected in a continuous 1.0 mol L-1 H2SO4 flow, and Ce(M is detected chronoamperometrically by reduction at the surface of a gold CDtrode. Such an electrode is made from a commercial CD that has a gold coating acting as a reflecting layer for the laser beam. It was found that a detection limit of 1.5 10(-7) Mol L-1 could be obtained with this method, that high concentrations (order of 10(-2) mol L-1) can be detected without IR drop and that the system can be used over several days without renewal or recalibration of the gold CDtrode. [References: 32]
机译:本文介绍了一种用于计时酸性溶液中Ce(IV)的计时安培测定的方法,以研究以Ce(IV)为引发剂的动力学和聚合反应机理。该方法基于FIA原理,其中将小样品以连续的1.0 mol L-1 H2SO4流量注入,并且通过在金CDtrode表面还原而在计时电流法中检测Ce(M)。市售CD具有镀金层作为激光束的反射层,发现该方法可获得的检出限为1.5 10(-7)Mol L-1,高浓度(10( -2)mol L-1)可以检测到而没有IR下降,并且该系统可以使用几天,而无需对CDtrode进行更新或重新校准。[参考文献:32]