首页> 外文期刊>Anaesthesia: Journal of the Association of Anaesthetists of Great Britain and Ireland >A prototype axial ultrasound needle guide to reduce epidural bone contact
【24h】

A prototype axial ultrasound needle guide to reduce epidural bone contact

机译:轴向超声针引导器原型,可减少硬膜外骨接触

获取原文
获取原文并翻译 | 示例
           

摘要

We have developed an ultrasound probe through the centre of which an epidural needle can pass, intended to reduce the rate of contact between bone and needle during epidural insertion. We tested the ability of this probe to identify the lumbar interspace, using A-mode ultrasound, in a submerged plastic model, a porcine phantom and five human volunteers. In the plastic model, the minimum echo representing the interspace was only 8.8% of the maximum echo from the 'bone'. In the porcine model, the echo variations between the interspace and L3 were up to 48% and the needle was safely inserted into the interspace without bone contact under guidance. The human study also showed that the maximum bone echo was at least three times stronger than the interspace echo. Axial ultrasound guidance, with the needle passing through the probe, offers a method for reducing bone contact during epidural insertion.
机译:我们已经开发了一种超声探头,硬膜外针可以穿过该探头,以减少硬膜外插入过程中骨头与针之间的接触率。我们在水下塑料模型,猪体模和五名人类志愿者中使用A型超声测试了该探头识别腰椎间隙的能力。在塑性模型中,代表空隙的最小回波仅为“骨头”最大回波的8.8%。在猪模型中,间隙和L3之间的回声变化高达48%,并且在引导下将针头安全地插入间隙中而没有骨骼接触。人体研究还表明,最大的骨回声至少比间隙回声强三倍。通过针头穿过探头的轴向超声引导提供了一种减少硬膜外插入过程中骨骼接触的方法。

著录项