Evaluation of agreement between two instruments in measurements of radiation dose rates in cats that underwent iodine I 131 treatment

机译:两种仪器在碘I 131治疗猫的辐射剂量率测量中的一致性评估

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

OBJECTIVE: To assess agreement between a commercially available Geiger-Meuller (GM) survey meter and millirem tissue-equivalent (TE) meter for measuring radioactivity in cats treated with sodium iodine I 131 ((131)I). ANIMALS: 15 cats with hyperthyroidism and undergoing (131)I treatment. PROCEDURES: Duplicate measurements were obtained at a distance of 30 cm from the thyroid region of each cat's neck by 2 observers who used both meters on day 3 or 5 after(131)I administration. Comparisons of measurements between meters and observers were made, with limits of agreement defined as the mean difference +/- 2 SDs of the differences. RESULTS: For observer 1, the mean of the differences in the 2 meters' measurements in all cats was 0.012 mSv/h (SD, 0.011 mSv/h). For observer 2, the mean of the differences in measurements was 0.012 mSv/h (SD, 0.010 mSv/h). For the GM meter, the mean of the differences of the 2 observers for all cats was 0.003 mSv/h (SD, 0.011 mSv/h). For the TE meter, the mean of the differences of the 2 observers for all cats was 0.003 mSv/h (SD, 0.007 mSv/h). CONCLUSIONS AND CLINICAL RELEVANCE: Results indicate that there was considerable agreement between meters and observers in measurements of radioactivity in cats treated with (131)I. Measurements obtained by use of the GM meter may be approximately 0.01 mSv/h less than or 0.03 mSv/h higher than those obtained with the TE meter. If this range is acceptable for an institution's release criteria, the 2 meters should be considered interchangeable and acceptable for clinical use.
机译:目的:评估市售的Geiger-Meuller(GM)测量仪和千里姆组织当量(TE)测量仪之间的一致性,以测量用碘I 131((131)I)处理的猫的放射性。动物:15例甲亢的猫正在接受(131)I治疗。程序:由两名观察员在距(131)I给药后第3天或第5天使用两米的两名观察员在距每只猫脖子的甲状腺区域30 cm处重复测量。比较了仪表和观察者之间的测量结果,一致极限定义为差的平均差+/- 2 SD。结果:对于观察者1,所有猫的2米测量值的平均值差为0.012 mSv / h(标准差为0.011 mSv / h)。对于观察者2,测量差异的平均值为0.012 mSv / h(SD,0.010 mSv / h)。对于GM仪表,所有猫的2个观察者的差异平均值为0.003 mSv / h(标准差为0.011 mSv / h)。对于TE仪表,所有猫的2个观察者差异的平均值为0.003 mSv / h(标准差为0.007 mSv / h)。结论和临床意义:结果表明,在用(131)I治疗的猫的放射性测量中,仪表和观察者之间有相当大的一致性。通过使用GM仪表获得的测量值可以比使用TE仪表获得的测量值低约0.01 mSv / h或高出0.03 mSv / h。如果该范围对于机构的发布标准而言是可接受的,则应将2米视为可互换并在临床上可接受。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号