首页> 外文期刊>Journal of Medical Imaging and Health Informatics >Ultra-Low Radiation Dose and Contrast Medium of Coronary CT Angiography Using Prospectively ECG-Triggered Without Heart Rate Control: A New Strategy
【24h】

Ultra-Low Radiation Dose and Contrast Medium of Coronary CT Angiography Using Prospectively ECG-Triggered Without Heart Rate Control: A New Strategy

机译:冠状动脉CT血管造影的超低辐射剂量和对比介质使用未经心率控制的前瞻性ECG触发:新策略

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The high pitch sprial (HPS) mode of Coronary CT angiography (CCTA) allows the radiation dose and contrast medium to be very low. However, the unsuccessful rate was to be considered, especially on high heart rate. The purpose of this study was to evaluate a new strategy without heart rate control by HPS and/ or Sequence mode in patients. A total of 46 patients were included (median age: 62.00 years old). At low heart rate (= 70 bpm), HPS mode with tube voltage 70-80 kVp and 30-40 mL contrast medium (270 mgI/ mL) for patients; Sequence mode with acquisition window (35%-45% of the R-R interval), tube voltage 70-80 kVp and 40-50 mL contrast medium for patients at high heart rate. A routine Sequence mode with wide acquisition window (30%-80% of the R-R interval) would be repeated when the coronary artery could not be assessed adequately. The image quality, radiation dose and contrast medium would be evaluated. In result, there were 3 patients with a repeated scanning. The effective dose (ED) was 0.51 +/- 0.41 mSv. The mean contrast medium and total iodine content were 39.78 +/- 12.91 mL and 10741.30 +/- 3485.18 mg. The assessment rate of coronary artery segments was 100%. In conclusion, different combinations of HPS and/or Sequence mode could provide satisfactory image assessment and minimize the radiation dose and contrast medium without heart rate control.
机译:冠状动脉CT血管造影(CCTA)的高间距升级(HPS)模式允许辐射剂量和造影剂非常低。然而,要考虑不成功的率,特别是在高心率。本研究的目的是通过HPS和/或患者序列模式评估没有心率控制的新策略。共有46名患者(中位年龄:62.00岁)。在低心率(& = 70bpm),HPS模式具有管电压70-80 kVp和30-40mL造影剂(270 mgi / ml),适用于患者;序列模式具有采集窗口(R-R间隔的35%-45%),管电压为70-80 kVp和40-50mL患者的患者,以高心率。当冠状动脉无法充分评估冠状动脉时,将重复具有宽采集窗口的常规序列模式(r-r间隔的30%-80%)。将评估图像质量,辐射剂量和造影介质。结果,有3例患者重复扫描。有效剂量(ED)为0.51 +/- 0.41msV。平均造影剂和总碘含量为39.78 +/- 12.91ml和10741.30 +/- 3485.18 mg。冠状动脉段的评估率为100%。总之,HPS和/或序列模式的不同组合可以提供令人满意的图像评估,并最小化无需心率控制的辐射剂量和造影剂。

著录项

  • 来源
  • 作者单位

    Inner Mongolia Med Univ Affiliated Hosp Dept Radiol Hohhot 010050 Inner Mongolia Peoples R China;

    Inner Mongolia Med Univ Affiliated Hosp Dept Radiol Hohhot 010050 Inner Mongolia Peoples R China;

    Inner Mongolia Med Univ Affiliated Hosp Dept Radiol Hohhot 010050 Inner Mongolia Peoples R China;

    GE Healthcare Dept Adv Applicat Shanghai 201203 Peoples R China;

    Inner Mongolia Med Univ Affiliated Hosp Dept Radiol Hohhot 010050 Inner Mongolia Peoples R China;

    Inner Mongolia Med Univ Affiliated Hosp Dept Radiol Hohhot 010050 Inner Mongolia Peoples R China;

    Inner Mongolia Med Univ Affiliated Hosp Dept Radiol Hohhot 010050 Inner Mongolia Peoples R China;

    Inner Mongolia Med Univ Affiliated Hosp Dept Radiol Hohhot 010050 Inner Mongolia Peoples R China;

    Inner Mongolia Med Univ Affiliated Hosp Dept Radiol Hohhot 010050 Inner Mongolia Peoples R China;

    Inner Mongolia Med Univ Affiliated Hosp Dept Radiol Hohhot 010050 Inner Mongolia Peoples R China;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 放射卫生;
  • 关键词

    Radiation Dose; Image Quality; Coronary Angiography; Contrast Medium;

    机译:辐射剂量;图像质量;冠状动脉造影;对比介质;