Effect of Contact Area on the Wear and Friction of UHMWPE in Circular Translation Pin-on-Disk Tests

机译:接触面积对圆形翻译引脚磁盘试验中UHMWPE磨损和摩擦的影响

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Circular translation pin-on-disk (CTPOD) tests were performed for ultrahigh molecular weight polyethylene (UHMWPE) with a view to reproducing wear mechanisms that prevail in total hip prostheses. The contact surface diameter varied from 3.0mm to 30 mm, while the slide track diameter was fixed, 10 mm. The counterface was polished CoCr, and the lubricant was diluted alpha calf serum. Either the nominal contact pressure (1.1 MPa) or the load (126 N) was kept constant. With a constant contact pressure, the wear factor decreased steeply when the contact diameter exceeded the slide track diameter, apparently because the wear debris was not readily conveyed away from the contact. With constant load, both the wear factor and the coefficient of friction increased linearly with increasing contact area. This trend was in agreement with clinical observations that the wear rate of UHMWPE acetabular cups increased with increasing femoral head size. With nominal contact pressures approaching 10 MPa however, the bearing surface topography markedly differed from clinical observations. This was probably due to overheating of the contact and plastic deformation that resulted in the formation of protuberances not seen clinically. The present study emphasized the importance of test parameters in the pin-on-disk wear screening of prosthetic joint materials. It appeared that the contact surface diameter of the flat-on-flat test should be below the slide track diameter, and that the nominal contact pressure should be of the order of 1 MPa.
机译:对超高分子量聚乙烯(UHMWPE)进行圆形翻译引脚(CTPOD)试验,以复制在总髋关节假体中占上铅的磨损机制。接触表面直径从3.0mm到30 mm变化,而滑道直径为固定,10毫米。配对面积被抛光COCR,润滑剂稀释α小牛血清。标称接触压力(1.1MPa)或负载(126n)保持恒定。随着恒定接触压力,当接触直径超过滑动轨道直径时,磨损因子显然降低,显然是因为磨损碎片不容易远离接触。由于恒定负载,磨损因子和摩擦系数随着越来越多的接触区域而导致线性增加。这种趋势符合临床观察,即UHMWPE髋臼杯的磨损率随着股骨头尺寸的增加而增加。然而,具有10MPa的标称接触压力然而,轴承表面形貌与临床观察显着不同。这可能是由于接触和塑性变形的过热,导致临床上没有看到突起的形成。本研究强调了试验参数在假体关节材料的蜗杆磨损筛选中的重要性。似乎平板上的接触表面直径应低于滑轨直径,标称接触压力应为1MPa的顺序。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号