Comparative Evaluation of the DPX and DPX-IQ Lunar Densitometer Systems Following Software Upgrade.

机译:软件升级后DPX和DPX-IQ月球密度计系统的比较评估。

获取原文
获取原文并翻译 | 示例
       

摘要

Bone densitometry research departments perform system and software upgrades infrequently in order to maintain high precision. This study compares the results obtained on a Lunar densitometer with DPX, and DPX-IQ installed to achieve year 2000 compliance. The DPX-IQ provides an improved femur edge detection algorithm with an expanded reference database. Two hundred data files for each measurement site acquired on DPX were reanalyzed on DPX-IQ. There was no change to the bone mineral density (BMD), bone mineral content (BMC), T-scores or Z-scores for the L2-L4 spine, radius (ultradistal and 33%), and total body. There was a significant high correlation for the femoral neck BMD (r = 0.98; p < 0.05). The mean differences in BMD, BMC, T-scores, and Z-scores at the femoral neck and Ward's and trochanteric regions were not significant (p > 0.05). The limits of agreement within the 95% confidence interval for the femoral neck BMD using the Bland and Altman method was between -0.057 and 0.063 g/cm2. This order of magnitude magnifies the long-term precision error and alters the usual confidence limits for interpretation of true change in densitometry practice. Therefore, it is important for reanalysis of DPX data files with the DPX-IQ to be performed so that longitudinal changes in BMD can be accurately assessed.
机译:骨密度研究部门不经常进行系统和软件升级,以保持高精度。本研究比较了用DPX的月球致密度计获得的结果,并安装了DPX-IQ,以实现2000年的遵守情况。 DPX-IQ提供具有扩展参考数据库的改进的股骨边缘检测算法。在DPX上获取的每个测量站点的两百个数据文件在DPX-IQ上重新分析。骨矿物密度(BMD),骨矿物质含量(BMC),T分数或Z分数没有变化,用于L2-L4脊柱,半径(超频和33%)和总体。股骨颈BMD有显着高的相关性(r = 0.98; p <0.05)。 BMD,BMC,T分数和股病房和沿促脱色区域的平均差异和Z分数并不显着(P> 0.05)。使用平坦和替代方法的股骨颈部BMD的95%置信区间内的一致性限制在-057和0.063g / cm 2之间。此幅度级倍率放大了长期精度误差,并改变了通常的置信度限制,以解释了密度测定实践的真实变化。因此,对于要执行的DPX-IQ进行DPX数据文件的重新分析是重要的,以便可以准确地评估BMD中的纵向变化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号