首页> 外文期刊>Journal of clinical engineering >Examining Root Causes of Medical Device Alarm System Failures Using Reports From Food and Drug Administration's Medical Product Safety Network (MedSun)
【24h】

Examining Root Causes of Medical Device Alarm System Failures Using Reports From Food and Drug Administration's Medical Product Safety Network (MedSun)

机译:检查医疗器械报警系统的根原因使用食品和药物管理医疗产品安全网(Medsun)的报告

获取原文
获取原文并翻译 | 示例
       

摘要

This article analyzes medical device reports related to medical device alarm system performance issues by categorizing the underlying failure causes and examining alarm system misses and false alarms. When a miss or false alarm occurs, the typical causes of the failure were either device or operator related. The most typical device-related failure was an unexpected component failure, whereas the typical operator-related failure resulted from a human factor performance issue with the device. These results suggest that, by improving the quality and usability of medical devices, the number of problematic false alarms and misses may be reduced.
机译:本文通过对潜在的故障原因和检查报警系统未命中和误报来分析与医疗器械报警系统性能问题相关的医疗器械报告。 发生未命中或误报时,故障的典型原因是设备或操作员。 最典型的设备相关的故障是意想不到的分量故障,而典型的操作员相关的失败是由设备的人为因素性能问题产生的。 这些结果表明,通过提高医疗器械的质量和可用性,可能会降低有问题的误报和未命中的数量。